Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Influence of radial clearance on Tresca stress in Al2O2-on-Al2O3 bearings for total hip prosthesis evaluated using
Muhammad Imam Ammarullah1,2,3, Abdulfatah Abdu Yusuf4,5, Muhammad Kozin6
1Bone Biomechanics Laboratory (BBL), Department of Biomedical Engineering and Health Sciences, Faculty of Electrical Engineering, Universiti Teknologi Malaysia, Johor Bahru, 81310, Johor, Malaysia. imamammarullah@gmail.com.
Abstract:
Ceramic materials have gained prominence as advanced alternatives to metals and polymers in total hip prosthesis, owing to their superior wear resistance, chemical inertness, and minimal ion release. Among ceramic biomaterials, aluminum oxide (Al2O2) is particularly favored for its exceptional hardness, high compressive strength, and excellent tribological performance, outperforming other candidates such as zirconium dioxide (ZrO₂) and silicon nitride (Si2N4). However, the inherently brittle nature of ceramics necessitates careful stress analysis to mitigate fracture risks in ceramic-on-ceramic bearing systems. This study investigates the influence of radial clearance as a key geometric parameter on the mechanical safety of Al2O2-on-Al2O3 bearings by analyzing Tresca stress distributions under physiologically representative loading conditions. A 2D axisymmetric finite element model was developed to simulate the internal stress behavior of the bearing components during a full gait cycle. Six radial clearance values, ranging from 0.03 to 0.3 mm, were systematically evaluated to determine their effect on internal shear stress patterns. The simulation results demonstrate that a radial clearance of 0.03 mm yields the lowest Tresca stress, indicating a more favorable stress distribution and reduced risk of shear-induced fracture. Conversely, larger clearances were associated with elevated shear stresses, which could compromise the mechanical integrity of the ceramic components over time. These findings highlight the importance of precise control over radial clearance during the design and manufacturing of ceramic-on-ceramic hip implants. The study provides foundational insights for optimizing implant geometry to enhance the structural reliability and clinical longevity of ceramic-bearing total hip prostheses.
Related Concept Videos
Stress Concentrations in Circular Shafts
Residual Stresses in Circular Shafts
Design of Transmission Shafts - Stress Analysis
Bearings: Problem Solving
Bearing Stress
Due to the intricacy of these microforces, an average value, known as bearing stress, is often used by...
Plastic Deformation in Circular Shafts