Related Experiment Video
Updated: Jan 16, 2026

An Inertial Measurement Unit Based Method to Estimate Hip and Knee Joint Kinematics in Team Sport Athletes on the Field
Published on: May 26, 2020
Validation of Inertial Measurement Units for Measuring Lower-Extremity Kinematics During Squat-Pivot and Stoop-Twist
Rutuja A Kulkarni1,2, Rajit Banerjee1,3, Vicki Z Wang1
1Human Engineering Research Laboratories, VA Pittsburgh Healthcare System, Pittsburgh, PA 15206, USA.
Abstract:
Optokinetic motion capture (OMC) is the gold standard for measuring the kinematics associated with lifting posture. Unfortunately, limitations exist, including cost, portability, and marker occlusion. The purpose of this study is to evaluate the agreement between OMC and inertial measurement units (IMUs) for quantifying joint kinematics during squat-pivot and stoop-twist lifting tasks. Ten unimpaired adults wearing both IMUs and OMC markers performed 24 lifting trials. Correlation coefficients and Root Mean Square Error (RMSE) between IMU and OMC time-series signals were computed for trunk and lower-extremity joints. Peak values obtained from each system during each trial were analyzed via Bland-Altman plots. Results show high correlations for trunk, knee, and ankle flexion angles (>0.9) and ankle rotation angles (>0.7). Moderate correlation was found for trunk axial rotation and lateral flexion angles (0.5-0.7). RMSE was under 9° for each angle. Biases between systems ranged from 0.3° to 16°. Both systems were able to detect statistically significant differences in peak angles between the two postures (p < 0.05). IMUs show promise for recording field data on complex lifting tasks.