Cross-Scanner Harmonization of AI/DL Accelerated Quantitative Bi-Parametric Prostate MRI
Dariya Malyarenko1, Scott D Swanson1, Jacob Richardson1
1Department of Radiology, University of Michigan Medical School, Ann Arbor, MI 48109, USA.
Sensors (Basel, Switzerland)
|September 27, 2025
Summary
AI/DL-accelerated quantitative bi-parametric MRI (q-bpMRI) shows diagnostic quality comparable to standard MRI for prostate cancer. Bias correction harmonizes T2 and ADC values across vendors, enabling multi-vendor clinical use.
Area of Science:
- Medical Imaging
- Artificial Intelligence in Medicine
- Quantitative MRI
Background:
- Clinical adoption of AI/DL-aided quantitative bi-parametric MRI (q-bpMRI) requires cross-vendor validation and harmonization.
- Prostate cancer (PCa) diagnosis benefits from quantitative MRI metrics like T2 and apparent diffusion coefficient (ADC) mapping.
Purpose of the Study:
- To implement and validate an AI/DL-accelerated q-bpMRI protocol for prostate cancer.
- To assess image quality, quantitative accuracy, and harmonization potential across different MRI vendors.
Main Methods:
- An AI/DL-accelerated q-bpMRI protocol (5-echo T2, 4-b-value ADC) was tested on two vendors' 3T scanners in six PCa patients.
- Image quality was compared to standard-of-care (SOC) MRI using a Likert-like scale.
- Quantitative validation was performed using a multiparametric phantom with known T2 and diffusion kurtosis values.
Main Results:
- The 6-minute AI/DL q-bpMRI achieved diagnostic quality comparable to SOC.
- AI/DL showed improved signal-to-noise ratio (SNR) for ADC mapping compared to SOC.
- Acquisition parameter-dependent biases were identified for T2 and ADC, but were consistent across vendors after AI/DL reconstruction.
- Bias correction successfully harmonized lesion T2 and ADC values across different AI/DL acquisitions.
Conclusions:
- AI/DL-accelerated q-bpMRI is a viable tool for prostate cancer assessment, offering comparable diagnostic quality to SOC.
- The developed bias correction workflow enables harmonization of quantitative T2 and ADC mapping across multi-vendor platforms.
- This facilitates the clinical integration of AI/DL-enhanced quantitative MRI.