Ambient Oxygen Concentration Enrichment When Using a Face Mask Connected to the Breathing Circuit Versus a Venturi
Danielle B Horn1, Ken D Nguyen2, Joni M Maga3
1Pain Medicine, Jackson Memorial Hospital, Miami, USA.
None:
Background The anesthesiology standard of care requires limiting oxygen (O2) concentration to less than or equal to 30% during monitored anesthetic care for upper chest and head and neck surgery to reduce the fire risk. While Venturi devices are used to regulate O2 concentration, they generate high flow rates, which may increase O2 concentration near the face mask. Even small increases of O2 concentration greater than room air significantly increase combustibility, highlighting the need for careful O2 delivery strategies. Methods A face mask was attached to a full-body patient simulator. The mask was connected to 1) the breathing circuit via a 15-mm tracheal tube adapter using the anesthesia machine's O2 blender set at 30%, or 2) the auxiliary O2 port using a 31% Venturi device. Surgical drapes were placed on intravenous poles ("ether screen") to separate the anesthesia and surgical side. Inflow O2 flow was 6 L/min for both groups, with the Venturi delivering ≈52 L/min. After room air calibration, ambient O2 concentrations were measured at three sites on the anesthesia side and three on the surgical side of the drape at no flow and then 15, 30, and 60 minutes after O2 flow initiation. Experiments were repeated three times. Results The pooled differences in the O2 concentration increases between the two devices were 0.80% (95% CI 0.53 to 1.07, P<.001) on the surgical side and 2.22% (95% CI 2.02 to 2.42, P<.001) on the anesthesia side of the ether screen. At 60 minutes, the mean (SD) concentrations in the Venturi group were 22.99% (0.31%) on the anesthesia side and 21.72% (0.90%) on the surgical side. Corresponding values in the anesthesia circuit group were 21.07% (0.12%) and 21.00% (0.11%). Conclusions A 31% Venturi device resulted in clinically relevant increases in ambient O2 concentrations on both sides of the ether screen compared to using the anesthesia machine O2 blender set at 30%. The differences are substantive because each 1% increase in the ambient O2 concentration increases the combustion rate of cotton by 15%. The use of a Venturi device to regulate O2 concentrations in the presence of an ignition source during head and neck surgery is not recommended. We advocate for anesthesia machine manufacturers to provide an integrated O2 blender for the auxiliary gas outlet of all their products as a fire safety measure.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...