The impact of V̇-Com on chinstrap usage and study completion rates during CPAP titration
Ludovico Messineo1, Carmen Keasling2, Bryan Hughes2
1Division of Sleep and Circadian Disorders, Departments of Medicine and Neurology, Brigham & Women's Hospital & Harvard Medical School, Boston, Massachusetts.
Study Objectives:
Mouth leak during continuous positive airway pressure (CPAP) therapy is a common barrier to tolerability and adherence and is often managed with chinstraps. We investigated whether V̇-Com, a device that reduces inspiratory pressure by adding noncompensated resistance into the CPAP circuit, could lower the chinstrap usage rate during in-laboratory titration. Additionally, we tested the effect of V̇-Com on study abortion rate due to CPAP intolerance.
Methods:
We conducted a retrospective study of in-laboratory CPAP titrations performed at a multisite sleep center. Mouth leak was identified based on ventilator-reported leak (> 30 L/min), visual observation of mouth opening, or characteristic alterations in the flow signal. V̇-Com was added as a first-line intervention when mouth leak was detected. If V̇-Com proved ineffective, it was removed, and a chinstrap was utilized. Similarly, V̇-Com was used in the attempt to salvage studies that were otherwise going to be aborted due to positive airway pressure intolerance.
Results:
Among 1,632 titrations, 190 patients (12%) experienced mouth leak and 46 individuals requested to abort their titration due to positive airway pressure intolerance (3%). V̇-Com reduced the chinstrap usage rate by 68% and the study abortion rate by 91%. In sensitivity analysis modeling hypothetical placebo response rates, the number needed to treat to avoid chinstrap use in 1 additional participant remained as low as 5, even under conservative assumptions.
Conclusions:
V̇-Com was associated with reduced chinstrap use and fewer aborted studies during CPAP titration. This suggests that V̇-Com may serve as a useful tool to improve pressure tolerance and leak management.
Citation:
Messineo L, Keasling C, Hughes B, et al. The impact of V̇-Com on chinstrap usage and study completion rates during CPAP titration. J Clin Sleep Med. 2025;21(12):2121-2127.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation I: Indication and Settings