Evaluation of an automated antimicrobial susceptibility testing system performance for colistin susceptibility in
Arzu İlki1,2, Fatih Mehmet Akıllı1,2,3, Sevim Alpsoy Özsoy2
1Department of Medical Microbiology, Marmara University School of Medicine, Istanbul, Turkey.
Abstract:
This study aims to assess the efficacy of VITEK 2 AST-XN21 (bioMérieux, Marcy l'Étoile, France) cards in identifying colistin susceptibility, comparing their performance with broth microdilution (BMD) results. A total of 187 carbapenem-resistant Acinetobacter baumannii were included and identified by MALDI-TOF (VITEK MS, bioMérieux, Marcy l'Étoile, France), and colistin susceptibility was determined by BMD and the VITEK 2 AST-XN21 card. The cs02n version is used in the VITEK 2. Isolates were evaluated for essential (EA) and categorical (CA) agreements, error rates, and ISO 20776-2:2021 bias rate (acceptable if bias rate is between -30% ≤ bias ≤+30%). Minimum inhibitory concentration (MIC) was interpreted according to the European Committee on Antimicrobial Susceptibility Testing breakpoints (susceptible ≤2 µg/mL; resistant >2 µg/mL). All isolates were subjected to in-house multiplex PCR for possible resistance genes mcr-1, mcr-2, mcr-3, mcr-4, and mcr-5. Out of 187 isolates, 16.04% (n = 30) of all isolates were determined as colistin resistant by the BMD method. MIC50 and MIC90 of the isolates were detected as 1 and 4 mg/L, respectively. When VITEK 2 AST-XN21 results were compared with BMD, CA and EA were 94.6% (177/187) and 89.1% (156/175), respectively. Very major error (3.2%, 6/187) and major error (2.1%, 4/187) were detected. ISO 20776-2:2021 bias rate was 3.6%. No mcr1-5 genes were detected.
Importance:
Colistin resistance in carbapenem-resistant Acinetobacter baumannii (CR-Ab) remains a global problem, and broth microdilution constitutes the workload of routine clinical microbiology laboratories. This study is the first to evaluate the cs02n version in the VITEK 2. Our results suggest that VITEK 2 AST-XN21 cards could be an alternative method to detect colistin resistance in CR-Ab isolates.
More Related Videos
05:06Author Spotlight: Advancing Antibiotic Resistance Research Using an Efflux-Deficient Bacterial Strain and a Single-Copy Gene Expression System
Published on: January 5, 2024
07:59Rapid and Specific Detection of Acinetobacter baumannii Infections Using a Recombinase Polymerase Amplification/Cas12a-based System
Published on: April 25, 2025