Analytical performance and user-friendliness of cobas pulse glucose meter in the hands of the intended users
Christine Morken1, Joakim Hekland1, Johan Olausson2
1The Norwegian Organization for Quality Improvement of Laboratory Examinations (Noklus), Haraldsplass Deaconess Hospital, Bergen, Norway.
Abstract:
Diabetes is a growing global health concern affecting millions worldwide. Glucose measurement remains a cornerstone in the diagnosis, monitoring, and management of this disease. Considering that numerous point-of-care (POC) devices for glucose measurements are available, systematic evaluation of their analytical performance and user-friendliness is essential. This study presents the findings of an independent evaluation of the cobas pulse glucose meter (Roche Diagnostics GmbH) conducted by the Scandinavian evaluation of laboratory equipment for point of care testing (SKUP). The study, which assessed both the analytical performance and user-friendliness, included 217 participants with diabetes and was conducted in both a hospital laboratory and primary healthcare settings. Capillary whole blood measurements were compared to a plasma glucose hexokinase method at Vejle hospital, Denmark. Certified reference materials from the National Institute of Standards and Technology (NIST) were used to correct all sample results. Repeatability and accuracy were assessed against predefined analytical performance specifications, and user-friendliness was evaluated using a structured questionnaire developed by SKUP. The cobas pulse demonstrated good repeatability (CV: 1.8-4.0%) and fulfilled the specification for accuracy, with minimal bias (maximum +0.16 mmol/L in the 7-10 mmol/L range). Haematocrit had no significant effect within the tested range (5-70%). User-friendliness was rated satisfactory, attributed to intuitive operation, rapid analysis times, and effective quality control features. In conclusion, the cobas pulse glucose meter met the performance criteria for professional use in clinical and primary care settings, supporting its suitability in near-patient glucose monitoring.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse Assessment Sites
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...