Related Experiment Video
Updated: Jul 30, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Percutaneous closure of patent foramen ovale with complex anatomy using the Memopart™ occluder device
S Gili1, F Cannata1, M Muratori1
1Centro Cardiologico Monzino, IRCCS, Milan, Italy.
Abstract:
We conducted a single center observational study aimed to assess safety and efficacy of Balmed Memopart PFO occluder in treatment of complex PFO anatomies including at least one of the following: tunnel length ≥ 8 mm, atrial septal aneurysm, lipomatous hypertrophy of fossa ovalis rims (≥10 mm), cribriform or additional multiple small defects on the fossa ovalis, residual large Eustachian ridge or Chiari network. This analysis demonstrated Memopart PFO occluder device to be a safe and effective option for PFO closure, with good procedural success rate (100 %) and no significant safety issues (1 mild residual shunt at 24 h).