Purpose:
Assess patient adherence, diurnal intraocular pressure patterns, and correlation with in-office tonometry of home intraocular pressure monitoring (IOP).
Patients And Methods:
In this prospective cohort study, 75 patients with open-angle disease (150 eyes) were recruited and loaned the iCare HOME2 tonometer (ICH2) after in-person training and instructed to take 6 measurements per day for 10 ± 4 days. Statistical analysis of adherence, diurnal IOP patterns, and correlation with in-office tonometry was conducted.
Results:
Most patients had good adherence (65% of patients took ≥ 6 measurements per day, median 7.4 measurements/day). However, adherence was poor for a notable minority of patients (19% of patients took less than 2 measurements per day). Patients with low adherence reported difficulty obtaining measurements and desire for more training with the ICH2 device. ICH2 identified diurnal patterns of IOP fluctuation: IOP varied significantly over the 24-hour day (p < 0.0001), and early morning peak IOP at 3 AM decreased significantly throughout the day to minimum mean IOP at 10 PM (p < 0.0001). IOP variance was significantly greater outside of clinic hours (p < 0.0001) and 36% of patients had maximum IOP outside of clinic hours. In-office IOP measurements had significant positive correlation to the closest ICH2 measurement obtained by the patient in time for both Goldman applanation (r2 = 0.35, p < 0.0001) and iCare IC100 (r2 = 0.48, p < 0.0001).
Conclusion:
ICH2 is clinically useable and effective for detecting out-of-office IOP spikes, diurnal IOP patterns, and IOP variation.
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Hypertension III: Clinical Manifestations and Diagnostic Studies
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure