Related Experiment Video
Updated: Jan 16, 2026

Transradial Access Chemoembolization for Hepatocellular Carcinoma Patients
Published on: September 20, 2020
Efficacy of an X-Ray Protective Drape in Reducing Operator Radiation Dose in Transfemoral Arterial Liver
Lawrence Bonne1, Niki Bergans2,3, Keir McCutcheon4
1Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000, Louvain, Belgium. lawrence.bonne@uzleuven.be.
Purpose:
Interventional radiologists are among the healthcare professionals most exposed to ionizing scatter radiation. A 0.5 mm lead equivalent X-ray protective drape (MXPD, Mavig, Germany) was introduced to reduce operator radiation dose and has demonstrated efficacy during cardiac interventions. This study evaluates the efficacy of the MXPD during transfemoral liver interventions.
Materials & Methods:
A single-center randomized controlled study was conducted to compare operator radiation exposure during transfemoral liver interventions (n = 100) with and without the MXPD. Primary outcome was the ratio of the combined operator radiation dose values-measured at the chest, eyes and both hands and normalized to the respective dose area products - between MXPD and control groups. Secondary outcomes included the ratios in operator radiation doses measured for the individual dosimeter locations.
Results:
The MXPD group showed a significantly lower normalized radiation dose per DAP (mean reduction 0.31 µSv; 95% CI 0.083-0.536, p = 0.009), representing a 43% reduction. The effect of the drape was the highest for the whole-body dose measured above the lead apron (55% reduction, p = 0.10), the left hand finger dose (43%, p = 0.20) and the eye lens dose estimated at the left temple (63%, p = 0.10) and in between the eyes (47%, p = 0.10).
Conclusion:
The MXPD can effectively reduce operator radiation dose during transfemoral arterial liver interventions, offering a practical adjunct to standard protective measures.
Level Of Evidence:
Level 1b, Randomized Controlled Trial.