Comparison of Post-Operative Outcomes between IOP-20 and IOP-50 in Phacoemulsification with Active Fluidics System:
Chandrashekhar M Wavikar1, Mamta N Tanna1, Gargi C Wavikar2
1Department of Ophthalmology, Wavikars Eye Hospital and Wavikar Eye Institute, Thane, Maharashtra, India.
Purpose:
To compare the central corneal thickness, mitigation events, post-operative vision, and complications in patients who underwent cataract surgery using the CENTURION ® Vision System with the ACTIVE SENTRY ® (Alcon Industries, Inc.) system and were assigned to one of two groups: intraocular pressure (IOP) group 20 or 50.
Setting:
Wavikar Eye Institute, Thane, India.
Design:
A prospective comparative two arm, randomized, single blinded, interventional study.
Methods:
A total of 107 eyes were randomized to either the IOP-20 (n=55) or IOP-50 (n=52) group. All patients underwent phacoemulsification with an active fluidics system performed by a single surgeon; intraocular pressure (IOP) was set at either 20 or 50 mmHg.
Results:
The mean [SD] baseline CCT in the IOP-20 group was 511.9 [35.7] μm and it was 521.1 [34.8] μm in the IOP 50 group (p=0.18). It had changed to 530.6 [39.8] μm in the IOP-20 group and 539.5 [39.2] μm in the IOP 50 (p=0.24). The mean [SD] ECD at the end of day 90 was 2423.0 [204.6] cells/mm2 in the IOP-20 group and 2476.1 [236.0] cells/mm2 in the IOP-50 group (p=0.22). Approximately 22% of the eyes in the IOP-20 group and 77% of the eyes in the IOP-50 group required active surge mitigation; this difference was statistically significant (p<0.001).
Conclusion:
We found that the phacoemulsification with an active fluidics system handpiece at 20 mmHg provided a safe and stable environment for cataract surgery.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment