Realtime detection of spontaneous circulation in humans during cardiopulmonary resuscitation using a continuous
Guro Mæhlum Krüger1,2, Sunniva Gjerald Birkeli1, Øystein Myrlund Hansen3,4
1Clinic of Anesthesia and Intensive Care, St. Olavs Hospital, Trondheim University Hospital, Trondheim, Norway.
Background:
The resuscitation society calls for precision-guided cardiopulmonary resuscitation (CPR), as current methods are inaccurate and time-consuming. RescueDoppler, a novel hands-free Doppler ultrasound system, continuously measures carotid blood flow during CPR. This pilot study assessed its performance, safety, and ability to detect chest compression-generated blood flow, spontaneous circulation, and return of spontaneous circulation (ROSC).
Method:
We investigated RescueDoppler in adult cardiac arrest patients at two centres, in-hospital (IHCA) and out-of-hospital (OHCA). The cardiac arrest team placed the RescueDoppler probe over the left common carotid artery with a self-adhesive patch, collecting blinded data during CPR. Data were later interpreted and time-synchronized with defibrillator data using custom MATLAB® software.
Results:
RescueDoppler was used in 26 IHCA and 36 OHCA patients from October 2023 to September 2024. Carotid blood flow curves were analyzed in 36 patients and synchronized with defibrillator data in 30. The RescueDoppler identified blood flow velocities generated by chest compressions and detected spontaneous circulation during rhythm checks. ROSC was defined by the presence of systolic and diastolic blood flow. No adverse events were reported but there were 22 device deficiencies mostly related to the self-adhesive patch and connecting cable. The system is user-friendly and requires minimal training.
Conclusion:
Real-time detection of blood flow in the carotid artery with hands-free Doppler ultrasound during CPR is safe and feasible, although the fastening patch and host unit need optimization. The RescueDoppler system detects spontaneous circulation and ROSC during rhythm checks and ongoing chest compressions. Further research is required to confirm clinical relevance.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation I: Adult
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Cardiopulmonary Resuscitation III: AED Use
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation