Introduction:
Chronic hypercapnic respiratory failure often necessitates non-invasive ventilation (NIV) at home. Our study aimed to assess the static and dynamic performance of six modern NIV home ventilators and one intensive care unit (ICU) ventilator and to identify performance clusters among the devices.
Methods:
A two-compartment lung model was connected to seven NIV ventilators (Sefam Stan, Philips A40, Philips DreamStation, Resmed Lumis 150, Löwenstein PrismaVent 30C, Löwenstein PrismaVent 40 and BellaVista 1000) in pressure-support mode. Static and dynamic (triggering and pressurisation) performances were assessed through three distinct clinical phenotypes and four levels of unintentional leak. Clustering analysis was performed using K-means.
Results:
For each of the seven ventilators, 144 conditions were tested, and a total of 3024 cycles were analysed. Static and dynamic performances were good to excellent across home ventilators, significantly higher than the ICU ventilator. Clustering analysis identified three performance clusters. Cluster 1 (Sefam Stan and Philips A40) showed significantly more precise accuracy of inspiratory pressure than Cluster 2 (Philips DreamStation, Resmed Lumis 150, Löwenstein PrismaVent 30C and Löwenstein PrismaVent 40) and Cluster 3 (BellaVista 1000): mean error=4.3%±5.1% versus 8.5%±6.7% versus 10.6%±14.7% respectively, p<0.001. For the triggering delay, Cluster 1 displayed shorter delays than Cluster 2 and Cluster 3 (41±5 ms vs 58±11 ms vs 67±13 ms, respectively, p<0.001). For the pressurisation delay, Cluster 1 displayed shorter delays than Cluster 2 and Cluster 3 (42±6 ms vs 64±14 ms vs 87±14 ms, respectively, p<0.001). For the pressure-time product at 300 ms, Cluster 1 displayed higher area under the curve for the first 300 ms than Cluster 2 and Cluster 3 (2.1±1.1 cmH2O /s vs 1.6±0.8 cmH2O/s vs 1.3±1.0 cmH2O/s, respectively, p<0.001). Continuous unintentional leaks did not modify the pressurisation performances in Cluster 1 but altered them in Clusters 2 and 3.
Conclusion:
The six NIV home ventilators demonstrated superior performance compared with the tested ICU ventilator. The ventilators of Cluster 1 were identified as top performers in clustering analysis and compensated for unintentional continuous leaks.
More Related Videos
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...