Agreement and repeatability of ocular surface function using the S390L Firefly WDR slitlamp compared with Keratograph
Linlin Su1,2, Tingting Yu1,2, Jiayan Chen1
1Department of Clinical Research, He Eye Specialist Hospital, Shenyang, 110034, China.
Scientific Reports
|October 7, 2025
Summary
The S390L Firefly WDR slitlamp shows good repeatability and agreement with the Keratograph 5M for measuring tear film parameters. This novel device is a viable alternative for clinical tear film evaluation.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Optometry
Background:
- Dry eye disease diagnosis relies on accurate tear film parameter assessment.
- The Keratograph 5M is a standard device for measuring tear film parameters.
- The S390L Firefly WDR slitlamp is a novel device for ocular surface analysis.
Purpose of the Study:
- To evaluate the agreement and repeatability of the S390L Firefly WDR slitlamp compared to the Keratograph 5M.
- To assess the clinical utility of the S390L Firefly WDR slitlamp for tear film measurements.
Main Methods:
- Prospective study involving 44 subjects.
- Tear film parameters measured: first non-invasive tear break-up time (NIBUTf), average non-invasive tear break-up time (NIBUTav), and tear meniscus height (TMH).
- Statistical analysis included Bland-Altman plots for agreement and Intraclass Correlation Coefficient (ICC) for repeatability.
Main Results:
- No significant differences were found in paired comparisons of NIBUTf, NIBUTav, and TMH between the two devices.
- Bland-Altman analysis indicated good agreement for NIBUTf, NIBUTav, and TMH.
- High ICC values (0.89 for NIBUTf, 0.84 for NIBUTav, 0.98 for TMH) demonstrated excellent repeatability for the S390L Firefly WDR slitlamp.
Conclusions:
- The S390L Firefly WDR slitlamp demonstrates good repeatability and acceptable agreement with the Keratograph 5M for tear film parameter measurement.
- The S390L Firefly WDR slitlamp can be considered a reliable alternative to the Keratograph 5M in clinical settings.
- The device offers acceptable intrasession repeatability for the clinical evaluation of tear film parameters.