Ultrasound endoscopes and routine microbiological surveillance - another critical device in reprocessing
Anna Mueller1, Bhavya Chakrakodi1, Marie-Theres Meier1
1Department of Infectious Diseases and Hospital Epidemiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
Routine microbiological surveillance of gastrointestinal endoscopes is crucial. Ultrasound endoscopes showed higher contamination rates compared to other types, indicating a need for targeted reprocessing improvements.
Area of Science:
- Medical Microbiology
- Infection Control
- Gastroenterology
Background:
- Periodic microbiological surveillance of gastrointestinal endoscopes is recommended to ensure effective reprocessing.
- Maintaining endoscope hygiene is critical for patient safety and preventing healthcare-associated infections.
Purpose of the Study:
- To evaluate differences in gastrointestinal endoscope contamination based on endoscope type.
- To assess the effectiveness of current reprocessing standards in preventing microbial contamination.
Main Methods:
- Microbiological analysis of 437 samples from various gastrointestinal endoscopes after reprocessing.
- Incubation on sheep blood agar, identification via MALDI-TOF MS or other methods, and colony-forming unit (cfu) counting.
- Defined contamination based on specific microbial presence (≥20 cfu/20 mL or specific pathogens) regardless of cfu count.
Main Results:
- Overall contamination rate was 2.5% (11/437 samples).
- Significant variation in contamination by endoscope type: ultrasound endoscopes (20.0%), colonoscopes (1.4%), gastroscopes (0.5%), duodenoscopes (0%), and enteroscopes (0%) (P < 0.001).
- Ultrasound endoscopes exhibited the highest proportion of contaminated samples.
Conclusions:
- A small overall proportion of gastrointestinal endoscopes were contaminated.
- Significant differences in contamination rates exist among endoscope types.
- Ultrasound endoscopes demonstrated the poorest performance regarding contamination, suggesting a need for enhanced surveillance and reprocessing protocols.
More Related Videos
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Endoscopic Procedures III: Video Capsule Endoscopy
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Endoscopic Procedures II: Colonoscopy