Comparison between vibration-controlled transient elastography and point shear wave elastography for assessment of
Tatsuya Gotoh1, Takashi Kumada2,3, Sadanobu Ogawa1
1Department of Imaging Diagnosis, Ogaki Municipal Hospital, Ogaki, Japan.
Purpose:
This study prospectively compared the diagnostic accuracy of vibration-controlled transient elastography (VCTE) and point shear wave elastography (pSWE) in the assessment of liver fibrosis, versus a reference standard of magnetic resonance elastography (MRE).
Methods:
This study prospectively enrolled patients with chronic liver disease, who underwent pSWE, VCTE, and MRE. Fibrosis was staged based on liver stiffness (LS) values measured using MRE: F0 (<2.61 kPa), F1 (≥2.61 to <2.97 kPa), F2 (≥2.97 to <3.62 kPa), F3 (≥3.62 to <4.69 kPa), and F4 (≥4.69 kPa). Each modality was performed independently, and the results were blinded to minimize bias. Diagnostic performance was assessed using the Pearson correlation coefficient (CC), Lin concordance correlation coefficient (CCC), area under the receiver operating characteristic curve (AUROC), Obuchowski index, integrated discrimination improvement (IDI), and net reclassification improvement (NRI).
Results:
In total, 251 patients (median age, 64 years; 97 women) were evaluated. Both pSWE (CC, 0.838; CCC, 0.825) and VCTE (CC, 0.803; CCC, 0.784) demonstrated strong correlations with MRE, with no statistically significant differences. AUROC values for diagnosing fibrosis stage were comparable between pSWE and VCTE. Based on the Obuchowski index, pSWE provided closer agreement with MRE in detecting ≥F1, ≥F2, and ≥F4. Analyses of IDI and NRI also displayed significantly better agreement between pSWE and MRE in detecting ≥F1, ≥F2, and ≥F4 (NRI: P<0.001; IDI: F1 and F2, P<0.001; F4, P=0.002).
Conclusion:
pSWE demonstrated closer alignment than VCTE with LS values measured by MRE, suggesting the potential of pSWE for noninvasively assessing liver fibrosis.