Comparison of Two Precurved Cochlear Implant Electrode Arrays from the Same Manufacturer: Electrophysiological
Khalil Rafiqi1, Kristoffer Foldager Jørgensen2, Anders Britze2
1Department of Otorhinolaryngology, Head and Neck Surgery, Aarhus University Hospital, Aarhus, Denmark, kharaf@rm.dk.
Introduction:
This study aimed to investigate differences between two precurved cochlear implant electrode arrays - Contour Advance (CA) and Slim Modiolar Electrode (SME) - in adult users, with a focus on electrically evoked compound action potential (ECAP) thresholds, 1-year postoperative C-levels, and battery consumption.
Methods:
Data from 70 adult cochlear implant recipients (35 CA, 35 SME) were analyzed. Intra- and postoperative ECAP thresholds were obtained using Cochlear Custom Sound software. One-year C-levels were measured through psychophysical loudness scaling. Estimated battery runtimes were examined in relation to median C-levels. A linear mixed-effects model was used to assess differences across electrode positions and between arrays. Model validation was performed through examination of standardized residuals versus fitted values, QQ-plots analysis, and comparison of observed versus expected serial correlations.
Results:
The CA array showed a more linear and more directly proportional relationship between ECAPs and C-levels. The SME array demonstrated significantly higher ECAPs and C-levels for the basal electrodes, larger pre- to postoperative changes in thresholds, and larger differences between ECAPs and C-levels, compared to the CA array. Battery runtime was significantly correlated with C-levels in the CA group, whereas impedance appeared more closely related to battery consumption in the SME group.
Conclusion:
Electrode array design affects electrophysiological and behavioral responses as well as battery consumption. We found significant differences between two precurved electrode arrays from the same manufacturer. These findings highlight the importance of array-specific programming strategies and suggest that electrode characteristics should be considered in clinical decision-making and device optimization.
More Related Videos
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
03:58Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Related Concept Videos
Comparison between RL and RC circuits
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Capacitors
When a voltage source is connected to a capacitor, positive and negative charges accumulate on the opposite plates. This accumulation generates a potential difference that equals the product of the...