Related Experiment Video
Updated: Jan 15, 2026

Closure of a Patent Foramen Ovale PFO: An Intervention Sequence
Published on: December 23, 2022
Transcatheter Closure of Patent Foramen Ovale With a Novel Biodegradable Device: A Prospective, Multicenter,
Fengwen Zhang1, Jie Dong1, Peijian Wei1
1Department of Structural Heart Disease, Fuwai Hospital & National Center for Cardiovascular Disease, Key Laboratory of Innovative Cardiovascular Devices, Chinese Academy of Medical Sciences & Peking Union Medical College, National Clinical Research Center for Cardiovascular Diseases, Beijing, China (F.Z., J.D., P.W., C.W., S.W., X.P.).
Background:
Novel biodegradable patent foramen ovale (PFO) closure devices offer a promising therapeutic option. The efficacy and safety of the novel biodegradable devices compared with nitinol devices have not yet been investigated in a randomized clinical trial.
Methods:
This multicenter, randomized, noninferiority trial examined whether the novel biodegradable PFO closure device achieves comparable closure success rates as traditional nitinol devices while demonstrating disappearance of the protruding parts of the device on echocardiography. A total of 190 patients with PFO were enrolled and randomly assigned to receive either the biodegradable device (n=96) or the nitinol device (n=94). The primary efficacy end point was PFO closure success rate at 6 months postprocedure as demonstrated by contrast echocardiography. We continued to evaluate device-related complications, device disappearance on echocardiography, and closure success rates over a 24-month follow-up period.
Results:
Successful PFO closure was achieved in 87 patients (90.63%) in the biodegradable device group and 86 patients (91.49%) in the nitinol device control group. The lower limit of the 95% CI of absolute difference was -8.98%, greater than the predefined noninferiority margin of -10%, confirming that the biodegradable device was not inferior to the nitinol device in terms of closure success. One patient in the trial group required surgical device removal because of intraprocedural deformation. No deaths, embolism, thrombus on the device, or erosion were observed in either group throughout the entire study period. Transthoracic echocardiography revealed that the hyperechoic area corresponding to the biodegradable device began decreasing within the first year after implantation and disappeared on echocardiography by 24 months after implantation.
Conclusions:
The novel biodegradable PFO closure device, which disappears on echocardiography within 24 months after implantation, demonstrates noninferiority to the traditional nitinol device in both efficacy and safety.