Related Experiment Video
Updated: Jan 15, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Reusable versus disposable forced air warming: A comparative study of thermal performance as a first step toward more
Jawad Matanis1, Amit Lehavi1, Aeyal Raz1
1From the Department of Anesthesiology, Rambam Healthcare Campus (JM, AL, AR, RV, ED, EM), Ruth and Bruce Rappaport Faculty of Medicine, Technion (JM, AL, AR, RV, ED, EM) and Nursing Department, Rambam Healthcare Campus, Haifa, Israel (EPB, BR).
Background:
Perioperative hypothermia increases the risk of surgical complications making effective patient warming essential. Disposable forced air warming blankets, while effective, contribute significantly to healthcare waste and environmental impact. Reusable alternatives may offer similar efficacy while reducing environmental footprint, but their thermal performance requires rigorous evaluation.
Objective:
To assess the safety and effectiveness of reusable textile blankets as an alternative to disposable forced air warming blankets, as a first step toward developing more sustainable perioperative practices.
Design:
We conducted an in-vitro, prospective, head-to-head comparison of heating performance in full-body and upper body designs, using three blanket types in a simulated operating-room scenario. Blanket types included disposable single use (DSU), light reusable cotton (LRC) and heavy reusable cotton (HRC). A single forced-air warmer was set to 38 °C for all tests, which simulated both normothermic (37 °C) and hypothermic (25 °C) patient conditions, measuring surface temperatures at six anatomically relevant points across the simulated body.
Setting:
Laboratory-based simulation in a controlled environment simulating operating room conditions with ambient temperature maintained at 20 ± 0.5 °C.
Participants:
A standardised simulated patient thermal model set to either normothermic (37 °C) or hypothermic (25 °C) conditions.
Main Outcome Measure:
Surface temperature and temperature homogeneity were measured at six anatomically relevant points across the simulated body.
Results:
The heavy cotton blanket provided the highest and most homogenous surface temperature in full-body designs (36.7 ± 0.6 °C) (DSU = 35.4 ± 1.1 °C; LRC = 35.7 ± 1.7 °C) (P < 0.05) and upper body designs (35.5 ± 0.5 °C) (DSU = 35.4 ± 0.5 °C; LRC = 34.9 ± 2.4 °C) (P < 0.05). Unsafe temperature elevations were not recorded for any of the test configurations. The statistical analysis revealed significant differences between blanket types (P < 0.05), with heavy reusable cotton consistently outperforming both disposable single-use and light-reusable cotton options in terms of temperature distribution.
Conclusion:
Carefully designed reusable blankets may provide a safe and environmentally friendly alternative to disposable forced air warming blankets, potentially reducing costs and environmental impact. Additional research examining infection control, durability through multiple laundering cycles and comprehensive life cycle analysis would further validate these promising findings.
Related Concept Videos
Mechanism of heat transfer
Mechanisms of Heat Transfer
Conduction, accounting for approximately 3% of body heat loss at rest, is the process of exchanging heat between molecules of two materials in direct contact. This can result in both heat loss and gain. For instance, when the body is submerged in water, which conducts heat 20 times more effectively than air, it can either lose or gain significant...
Mechanisms of Heat Transfer II
Mechanisms of Heat Transfer I
Thermal Stress
Decreased Body Temperature