Performance properties of filter-paper used in blood spot collection devices for quantitation of phenylalanine
Annabel Rodham1, Christopher von Ruhland2, Rachel Carling3,4
1Wales Newborn Screening Laboratory, Department of Medical Biochemistry, Immunology & Toxicology, University Hospital of Wales, Cardiff, UK.
Aims:
Accurate and precise measurement of dried blood spot (DBS) phenylalanine (Phe) is vital for managing phenylketonuria (PKU). Standard DBS collection devices use grade-226 filter-paper, while the CapitainerB quantitative device utilizes grade-222 filter-paper. Although grade-226 filter-paper performance is well characterized, data on grade-222 filter-paper are sparse. This study aimed to investigate the analytical properties of grade-222 and grade-226 filter-papers.
Materials And Methods:
We compared grade-222 and grade-226 filter-papers for Phe measurement accuracy and imprecision in DBS generated using both filter-papers. Scanning electron microscopy (SEM) and slit lamp imaging were used to assess the physical properties of the filter-papers.
Results:
Using an aqueous calibrator as reference, grade-222 exhibited a mean bias of -1.1%, the mean bias for grade-226 was -7.3%. Intra-assay imprecision was 2.3% for grade-222, versus 4.2% for grade-226. SEM revealed that fibers in grade-226 filter-paper are bonded by an amorphous material, which is absent in grade-222 filter-paper. Total error analysis indicated grade-222 filter-paper reduced uncertainty of Phe measurement compared to grade-226 filter-paper.
Conclusions:
Grade-222 filter-paper was proven to have superior analytical performance for Phe quantification, providing improved differentiation between safe and harmful Phe concentrations and offering more reliable PKU monitoring compared to traditional grade-226 filter-paper.