A 15-Min Method for Hydrophone Spatial Averaging Correction in Transcranial Neuromodulation System Characterization
1Center for Devices and Radiological Health, US Food and Drug Administration, Silver Spring, MD, USA.
Hydrophone spatial averaging significantly impacts transcranial ultrasonic stimulation (TUS) system characterization. This study provides a simple method to correct for these errors, improving TUS accuracy.
Area of Science:
- Neuroscience
- Biomedical Engineering
- Acoustics
Background:
- Transcranial ultrasonic stimulation (TUS) is a non-invasive neuromodulation technique.
- Accurate characterization of TUS systems is crucial for reproducible research and clinical applications.
- Hydrophone spatial averaging is a recognized potential source of error in acoustic measurements.
Purpose of the Study:
- To quantify the error introduced by hydrophone spatial averaging in TUS system characterization.
- To provide a practical method for researchers to correct these measurement errors.
- To emphasize the importance of accounting for spatial averaging in TUS system specifications.
Main Methods:
- Analysis of 74 TUS system specification sets from 60 peer-reviewed publications.
- Exclusion of studies using hydrophones larger than the International Transcranial Ultrasonic Stimulation Safety and Standards Consortium (ITRUSST) recommended maximum.
- Computation of spatial averaging error correction factors for acoustic intensity.
Main Results:
- Hydrophone spatial averaging introduces significant errors in TUS system characterization.
- Correction factors for acoustic intensity were calculated: 1.10 ± 0.07 for human parameters, 1.12 ± 0.14 for animal parameters, and 1.18 ± 0.17 for high-frequency animal parameters.
- These findings confirm spatial averaging as a consistent source of error.
Conclusions:
- Hydrophone spatial averaging is a significant and common source of error in TUS system characterization.
- A straightforward method is presented for researchers to correct these errors using their TUS system specifications.
- Implementing this correction method enhances the reliability and accuracy of TUS research.