Inter-institutional variability in CT-to-mass-density conversion tables for helical tomotherapy: a national survey in
Shogo Tsunemine1, Shuichi Ozawa2, Ryosei Nakada1
1Department of Radiation and Proton Office, Shizuoka Cancer Center, 1007 Shimonagakubo, Nagaizumi-cho, Sunto-gun, Shizuoka 411-8777, Japan.
None:
This study evaluates current practices and challenges associated with computed tomography number-to-mass density (CT-MD) conversion tables in helical tomotherapy across Japan and explores directions for standardization and quality improvement amid the increasing adoption of adaptive radiotherapy (ART). A nationwide web-based survey was conducted across 34 institutions utilizing the Radixact system. Data were collected on CT acquisition protocols, calibration phantoms, density plugs, reconstruction algorithms, table registration timing and quality assurance (QA) frequency. Registered CT-MD tables were categorized by CT modality: Simulation CT (SimCT), ClearRT and CTrue. ClearRT tables were analyzed by phantom setup (full vs half), and CTrue tables by reconstruction method [filtered back projection (FPB) vs iterative reconstruction (IR)]. Inter-institutional variations in CT numbers and the number of data points were assessed. SimCT tables exhibited the widest variation in the number of data points (median = 10) and high-density CT numbers. ClearRT tables (median = 8) showed variations of up to 300 Hounsfield units (HU) in cortical bone; the half-phantom setup reduced inter-institutional variability. CTrue tables (median = 8) demonstrated high consistency, with negligible differences between IR and FPB. All plug CT numbers of CTrue remained within the tolerance defined by the American Association of Physicists in Medicine Task Group 148. However, CT numbers for air plugs varied by ~±30 HU, indicating inconsistent handling of air reference values. Additionally, 43% of institutions did not perform routine QA. Standardizing phantom geometry, air CT number handling and QA protocols-particularly using half-phantom calibration-may improve CT-MD table consistency and dose accuracy in ART.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography