Related Experiment Video
Updated: Jan 15, 2026

Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
Flow monitoring of venovenous and venoarterial ECMO to detect circuit obstructions using hemodynamic modeling
Ruth White1, Srivats Sarathy2, Aditya Badheka3
1Department of Biomedical Engineering and Science, Florida Institute of Technology, Melbourne, FL, USA.
Abstract:
IntroductionCurrent extracorporeal membrane oxygenation (ECMO) monitoring methods for circuit obstructions have limitations. In this study, we test the hypothesis that changes in flow in a pre-existing shunt tubing of the ECMO circuit can be used as a predictor of obstructions for both venovenous (V-V) and venoarterial ECMO (V-A ECMO) circuits using a computational hemodynamic model.MethodsA custom computational hemodynamic model (CHM) representing the blood flow within the V-A ECMO and V-V ECMO circuit and native cardiovascular system was developed for a hemodynamic parametric analysis. Several degrees of obstructions within the oxygenator and outlet cannula were simulated, and the flow in the ECMO circuit and shunt was monitored to assess its sensitivity to circuit obstructions. Over 500 cases were analyzed to investigate various clinically-relevant levels of ECMO support (2 - 5.5 L/min).ResultsOur model demonstrated that shunt flow changed linearly with increasing oxygenator obstruction for all oxygenator obstruction scenarios tested, which matched experimental data. Similar relationships were also seen for all outflow cannula obstructions.ConclusionsThe findings are consistent with our hypothesis, indicating the sensitivity of the shunt flow rate as a predictor of ECMO circuit obstructions for both V-A and V-V ECMO circuits. The data supports that shunt flow responds to hemodynamic changes throughout the ECMO-patient circuit and thus can be used as an early indicator for ECMO system obstructions regardless of V-A or V-V ECMO setup, indicating its robustness as an early warning indicator of ECMO system obstructions.
Related Concept Videos
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...