Introduction:
The ISO 15189:2022 standard considers both the robustness of analytical methods and the risk of erroneous results in the quality control plan (QCP) design. Westgard et al.'s nomogram recommends quality control (QC) rules based on sample run size to ensure that the maximum expected number of unreliable patient results remains below one. This study aimed to implement a standardized, risk-based QC strategy across multiple analyzers without integrated on board QC, ensuring practical quality assurance.
Material And Methods:
Thirty-two biochemistry parameters on Alinity c systems and three on Cobas Pro systems were included. The analytical performance of each parameter on each analyzer was assessed using sigma metric. Workload requirements were considered to determine the desired run size. Based on the "sigma metric statistical QC run size nomogram" proposed by Westgard et al., a multistage bracketed QCP was designed for each parameter. When multiple designs were available, the simplest QC rule was prioritized.
Results:
Seven QCPs were initially established for 35 parameters. In the absence of automation, practical adaptations based on sigma metrics were implemented. Additionally, to streamline management, the QCP covering the greatest number of parameters per analyzer was prioritized, which ultimately resulted in the adoption of only two general QCP. Only 4 individualized QCP were required to cover 10 parameters with lower sigma values.
Conclusions:
This approach demonstrates the feasibility of implementing a refined QC strategy for parameters with sigma ≥ 4 in a highly automated laboratory, ensuring consistent quality assurance and efficient resource allocation for higher-risk tests.
More Related Videos
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
08:58Efficient Sampling of Genetically Encoded Biosensor Design Space Enabled with a Design of Experiments and Automation Workflow
Published on: October 17, 2025
Related Concept Videos
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Introduction to Statistical Process Control
Quality Assurance
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Data Validation
Key parameters for method validation include:
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...