Simultaneous dual-isotope SPECT imaging using a cardiac cadmium-zinc-telluride camera with various Technetium-99 m to
Takayuki Shibutani1, Masahisa Onoguchi2, Hiroto Yoneyama3
1Department of Quantum Medical Technology, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80, Kodatsuno, Kanazawa, Ishikawa, Japan.
Objective:
Cadmium-zinc-telluride (CZT) SPECT cameras offer high energy resolution. This allows simultaneous dual-isotope (SDI) acquisition even when 123I and 99mTc photon peaks are close. This study aimed to reveal the impact of image quality for 99mTc/123I ratios on SDI images.
Methods:
We created normal and inferior wall defect myocardial models using an anthropomorphic myocardial phantom. The doses of 99mTc/123I ratios were set to 1.0, 3.5, 5.0 and 6.5 (referred to as Tc/I_1.0, Tc/I_3.5, Tc/I_5.0 and Tc/I_6.5 conditions). We acquired SPECT images using D-SPECT Cardio. The acquisition time was adjusted to achieve 1.5 × 106 left ventricular 123I counts as a clinical reference. Short-axis images were reconstructed with (SC) and without (NC) scatter correction. Kernel and Gaussian-standard values were set to 1 and 1.0 as default conditions. The filter parameters were changed to 1, 3 and 5 for the kernel settings, and 0.25-1.0 for Gaussian standards. The image quality of normal and defective myocardia at different 99mTc/123I ratios was evaluated as cavity contrast on short-axis images, the percent coefficient of variance (%CV) and %uptake on polar maps. In addition, these reconstruction conditions were applied to a patient with acute coronary syndrome, supporting the phantom-based findings.
Results:
In the normal myocardial phantom, %uptake in the inferior and inferolateral myocardial walls was significantly lower in SC images than in NC images for both 99mTc and 123I. In SC images, 99mTc counts in the inferoseptal myocardial wall slightly decreased along with the 99mTc/123I ratio, whereas 123I counts in the inferolateral walls declined as the 99mTc/123I ratio increased to 6.5. Higher kernel and Gaussian standard settings resulted in lower %CV values in both NC and SC 99mTc and 123I images, improving image uniformity. Overall, NC images showed more homogeneous myocardial distribution in the inferior wall compared with SC images.
Conclusion:
Image quality and correction effects are influenced by 99mTc/123I ratios, resulting in heterogeneity in the inferior wall of the normal myocardium due to SC. Adjusting kernel parameters of the Gaussian filter or using NC images may help improve image quality when 99mTc/123I ratios are very high or low.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...