Developing Low-Viscosity Lubricants to Maintain Lubricity between Overtube and Enteroscope for Improved
Katsuma Yamauchi1,2, Ryohei Hirose1,2, Taku Kanou1,2
1Department of Molecular Gastroenterology and Hepatology, Graduate School of Medical Science, Kyoto Prefectural University of Medicine, 465 Kajii-cho, Kawaramachi-Hirokoji, Kamigyo-ku, Kyoto 602-8566, Japan.
Abstract:
During enteroscopy, the hydrophilic coating on the inner surface of the overtube (OT) reduces the dynamic friction coefficient (DFC) between the endoscopic shaft (ES) and OT, allowing smooth insertion of the ES. However, endoscopists often encounter resistance to ES insertion and poor endoscopic manipulation during prolonged enteroscopy. We hypothesized that the reason is the degradation of the hydrophilic coating due to the continuous contact between the ES and OT and the associated increase in DFC. Focusing on single-balloon enteroscopy, we developed a unique model to continuously measure the DFC for 30 min while moving the ES in contact with the inner surface of the OT, which was applied with pure water or various lubricants. When pure water was applied, the initial DFC was 0.112 ± 0.001; however, it increased to 0.357 ± 0.002 over 30 min (219% increase). Meanwhile, the application of 0.1% carboxymethyl cellulose, 0.05% sodium alginate, 0.25% hydroxyethyl cellulose, and 0.04% xanthan gum solutions resulted in significantly lower initial DFCs of 0.073 ± 0.0004, 0.081 ± 0.0004, 0.079 ± 0.0004, and 0.063 ± 0.0003, respectively. Moreover, low DFCs were maintained even after 30 min, with DFC increase rates of +2.5%, -0.2%, -5.9%, and +5.6%, respectively. In addition, the injection pressures required for these optimal lubricants were extremely low. The optimal lubricants identified in this study successfully maintained lower DFCs than pure water, which is conventionally used to induce lubrication between the ES and OT; therefore, they may improve endoscopic manipulation.
Related Concept Videos
Drugs Affecting GI Tract Motility: Other Laxatives
Osmotic or saline laxatives, like magnesium hydroxide or milk of...
Design Example: Deciding Thickness of Lubricating Fluid in a Shaft
To calculate the required thickness of the lubricant layer, the tangential velocity at the shaft's surface must first be determined. This velocity is calculated by converting the rotational speed to angular velocity...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures III: Video Capsule Endoscopy