Related Experiment Video
Updated: Jan 14, 2026

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Complications Associated With Transvenous Cardiac Implantable Electronic Devices: Recognition and Management : A
Peter J Zimetbaum1, Enrico G Ferro1, Eric A Secemsky1
1Division of Cardiovascular Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston; and Richard A. and Susan F. Smith Center for Outcomes Research, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts (P.J.Z., E.G.F., E.A.S., D.B.K.).
Insights
Cardiac implantable electronic devices (CIEDs) are common, but complications like lead failure, infection, venous obstruction, and tricuspid regurgitation require prompt management. Early recognition and multidisciplinary care are key for optimal patient outcomes.
Area of Science:
- Cardiology
- Medical Devices
- Clinical Medicine
Background:
- Cardiac implantable electronic devices (CIEDs) significantly improve patient quality and duration of life.
- Exponential growth in CIED utilization necessitates comprehensive understanding of device-associated complications.
Purpose of the Study:
- To provide an updated review of the four most common CIED complications for general internal medicine physicians and hospitalists.
- To present the latest evidence for managing CIED complications and their expected outcomes.
Main Methods:
- Review of current literature on CIED complications.
- Focus on incidence, clinical manifestations, and management strategies for generator/lead failure, CIED infections, lead-related venous obstruction, and lead-related tricuspid regurgitation.
Main Results:
- Generator/lead failure incidence is <1% annually, managed conservatively.
- CIED infections (approx. 1% annually) have high mortality (10-20%) and often require lead extraction.
- Lead-related venous obstruction (5-30%) has variable management with often suboptimal results.
- Lead-related tricuspid regurgitation (20-30%) is an emerging area with need for evidence on lead extraction vs. jailing.
Conclusions:
- While leadless devices may reduce complications, optimal management of existing CIED issues requires a multidisciplinary team approach.
- Prompt recognition and treatment of CIED complications are crucial, guided by evolving evidence and specialized expertise.
Abstract:
Cardiac implantable electronic devices (CIEDs) have increased the quality and duration of life for millions of patients. As their utilization continues to grow exponentially, clinicians need to diagnose and manage several device-associated complications that may arise during the CIED life cycle. This primer is tailored for the general internal medicine physician or hospitalist, who will inevitably take care of patients with CIEDs, to provide a contemporary update on the incidence and clinical manifestations of the 4 most common CIED complications, with the latest evidence to guide clinical management and expected outcomes. Specifically, this review focuses on: 1) generator/lead failure, which has decreased to less than 1% per year due to continuous advances in manufacturing and programming and is mostly managed conservatively with close monitoring; 2) CIED infections, which, while also rare at approximately 1% per year, carry a short-term mortality of 10% to 20%, thus requiring immediate diagnosis and often urgent treatment with lead extraction because antibiotic treatment is rarely effective; and 3) lead-related venous obstruction, which has highly variable incidence (5% to 30%) and manifestations, ranging from arm swelling to superior vena cava syndrome. Management options range from anticoagulation to balloon venoplasty, often with unsatisfying results. This review will also focus on 4) lead-related tricuspid regurgitation, which affects 20% to 30% of CIEDs and has become an area of intense interest with the development of percutaneous tricuspid treatments-where evidence is urgently needed to inform the need for lead extraction versus jailing during tricuspid interventions. The progressive adoption of leadless devices may significantly reduce many of these complications. Nonetheless, optimal management requires input from a multidisciplinary team of electrophysiologists and imaging, heart failure, and structural interventional specialists-who should be able to recognize and treat each complication promptly based on a rapidly evolving evidence base.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Cardiac Catheterization IV: Nursing Management
Cardiomyopathy V: Interprofessional Care
Endocarditis IV: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview
Endocarditis III: Medical Management
Cardiomyopathy VI: Nursing Management