Comparison of Pulse Oximetry and Blood Pressure Measurements Between Sensor Locations in Children During Dental
Brianne N Cipich1, Juan F Yepes2, Stephen Wilson3
1MSD Candidate, Department of Pediatric Dentistry, School of Dentistry;, Indiana University, Indianapolis, Ind., USA.
Abstract:
Purpose: To compare oxygen saturation measurements when pulse oximeter sensors were simultaneously placed on the index finger and the second toe, and compare blood pressure recordings with non-invasive blood pressure (NIBP) cuffs simultaneously placed on the upper arm and the lower calf in children receiving general anesthesia. Methods: A total of 50 children, aged three to nine years, defined as ASA Class I or II receiving comprehensive dental treatment in the operating room with general anesthesia, were included. Pulse oximeter sensors were simultaneously placed on each subject's index finger and ipsilateral second toe. Non-invasive blood pressure cuffs were simultaneously placed on the upper arm and ipsilateral lower calf above the ankle. The same anesthesia regimen was used on all patients. Following induction of general anesthesia, measurements were simultaneously recorded every five minutes during the dental procedure. Results: Index finger oxygen saturation measurements were significantly lower than the second toe (P<0.001). Arm systolic blood pressure measurements were significantly lower than calf measurements (P<0.001). Arm diastolic blood pressure measurements were significantly lower than calf measurements (P<0.001). Conclusions: Discrepancy of vital sign measurements may be notable depending on the site from which measurements are obtained. If blood pressure is taken on the calf of a child in the Trendelenburg position, the reading will likely produce a higher systolic and diastolic measurement and should be interpreted with caution. A pulse oximeter placed on the first finger or second toe of healthy children can be regarded as interchangeable in a clinical setting.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...