A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
Evaluating the efficacy of disinfectant agents and application times for vascular catheter needleless connector
Maryanne Kuek1, Sarah K McLean1, Enzo A Palombo1
1School of Science, Computing and Emerging Technologies, Swinburne University of Technology, Hawthorn, Victoria, Australia.
Background:
Needleless connectors (NCs) are attached to vascular catheters to provide an injectable closed circuit without risk of needlestick injuries but some have been associated with catheter-associated bloodstream infections (CABSIs), which increase treatment costs and hospital stays. Effective NC disinfection before catheter use is therefore critical. This study evaluated the efficacy of three disinfectant agents across varying decontamination durations and drying times.
Methods:
NC septa were inoculated with clinically isolated Acinetobacter courvalinii and Staphylococcus hominis from venous access devices (VAD) insertion sites and infusion tubing, then treated with either 70 % isopropyl alcohol (IPA), 2 % chlorhexidine gluconate in 70 % IPA (CHG-IPA) for 5, 10, or 15 s, or alcohol-impregnated antiseptic caps (AICs) for 5 min. Treatments were followed by drying times of 5, 10, or 15 s. Bacterial recovery was assessed via vortexing and sonication in Dey-Engley neutralising broth, followed by spread-plating on Tryptic Soy Agar and incubation at 37 °C. A total of 378 decontamination procedures were performed, with nine replicates per condition.
Results:
All disinfectant treatments significantly reduced (p < 0.05) bacterial counts compared to positive controls, achieving reductions of approximately 2.4-4.4 log CFU/mL. However, no statistically significant differences (p > 0.05) were observed among disinfectant types, application durations, or drying times.
Conclusion:
NC disinfection proved essential in reducing bacterial counts. While no method proved superior, the study may have been underpowered for some comparisons and larger studies should further evaluate both CHG in IPA and AICs.
Related Concept Videos
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Chemical Agents for Microbial Control
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Antimicrobial Effectiveness
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...