Balloon dilation of the para-device space to alleviate severe flow restriction in an oversized pulmonary flow
Raymond N Haddad1, Hanan Faraji2, Ahmed Adel Hassan3
1Centre de Référence Malformations Cardiaques Congénitales Complexes-M3C, Hôpital Universitaire Necker-Enfants Malades, Assistance Publique-Hôpitaux de Paris, Paris, France. raymondhaddad@live.com.
Background:
Manually modified microvascular plugs (MVPs) to pulmonary flow restrictors (PFRs) are innovative devices for managing pulmonary overcirculation percutaneously, though achieving optimal restriction while maintaining device stability can be challenging.
Case Presentation:
We describe a full-term newborn with a double outlet right ventricle and a large malaligned subaortic ventricular septal defect, complicated by pulmonary overcirculation. At 24 days of life (2.4 kg, 50 cm), bilateral MVP-based PFRs were implanted, with an MVP-7Q placed in the right pulmonary artery (4.1 mm diameter) and an MVP-5Q in the left (3.8 mm diameter), after fenestrating the MVPs by removing one triangle with a surgical scalpel. Immediate severe flow restriction to the right lung was managed with dilation of the para-right PFR space using 2 mm and then 3 mm semi-compliant coronary balloons, increasing oxygen saturation from 60 to 87%. By the end of the procedure, the Qp: Qs ratio decreased from 2.3:1.0 to 1.1:1.0. At 12 weeks, the patient weighed 4.8 kg, with oxygen saturation at 92% and Doppler gradients of 70 mmHg on the left and 45 mmHg on the right. Nine months post-intervention, the patient underwent biventricular repair and PFR removal at 7.9 kg. A fibrotic endothelial reaction with wall thinning on the right required autologous patch repair. A two-week follow-up ultrasound showed no stenosis.
Conclusions:
Para-device space dilation with semi-compliant coronary balloons effectively relieves severe flow restriction in oversized PFRs, restoring flow while preserving device integrity. However, long-standing oversized PFRs cause intimal damage, necessitating patch repair, despite excellent surgical outcomes.
Clinical Trial Number:
Not applicable.
Related Concept Videos
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pipe Flowrate Measurement
The orifice meter is a simple,...