Related Experiment Video
Updated: Jan 14, 2026

Transcorporal Artificial Urinary Sphincter Cuff Placement in a Case Requiring Revision for Urethral Atrophy
Published on: June 16, 2022
Comparative Urodynamic Study in Cadaver of Urethral Pressure Profilometry Between the Artificial Urinary Sphincter
Aurélien Beaugerie1, Elliot Tokarski1, Anne Denormandie1,2
1Department of Urology, Pitié-Salpêtrière Academic Hospital (AP-HP), Paris, France.
Introduction And Objective:
The artificial urinary sphincter (AUS) is currently the gold standard treatment for stress urinary incontinence in men, and it's also a treatment option for women in Europe. UroActive is a new electronic device that offers remotely adjustable settings, including device pressure. This study aims to compare the range of Maximal Urethral Closure Pressures (MUCPs) covered by UroActive with those covered by the current AMS800 in male and female cadavers.
Methods:
Six cadavers (3 males and 3 females) were implanted with an occlusive cuff (OC) positioned around the bulbar urethra in men and around the bladder neck in women. A MUCP measurement was performed for each of the 3 different AMS800 Pressure-Regulating Balloons (PRBs): 51-60, 61-70, and 71-80 cmH2O, that were successively connected to the OC. The AMS800 PRB was then replaced by the UroActive Control Unit (CU), and MUCP measurements were performed at set pressures from 10 to 150 cmH2O.
Results:
UroActive device remained fully functional (wireless communication, calibration) throughout the study period. UroActive CU achieved MUCPs values that encompassed the entire range observed with the 3 different AMS800 PRBs. A strong positive correlation between set device pressures sent to UroActive CU and MUCPs was noted in both males (r2 = 0.984) and females (r2 = 0.948).
Conclusion:
The findings suggest that UroActive provides a wide adjustable range of urethral closure pressures, potentially offering an alternative to AMS800 for managing stress urinary incontinence in both men and women. Further clinical studies are necessary to confirm its safety and effectiveness in patients.
More Related Videos
06:46Author Spotlight: Enhanced Urodynamic Method for Precise Urine Measurement in Awake Mice with Neurogenic Bladder
Published on: June 7, 2024
10:07Cystometric and External Urethral Sphincter Measurements in Awake Rats with Implanted Catheter and Electrodes Allowing for Repeated Measurements
Published on: January 30, 2018
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Imaging Studies VI: Voiding Cystourethrography and Cystography
Anatomy of the Genitourinary System II: Bladder and Urethra