A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
WS2/CuO-based non-enzymatic sensor for the detection of glucose in sweat
Dujuan Li1, Mingzhu Yang1, Wenhui Li1
1Ministry of Education Engineering Research Center of Smart Microsensors and Microsystems, School of Electronics and Information, Hangzhou Dianzi University, Hangzhou, 310018, China.
Background:
Traditional blood glucose monitoring requires piercing the skin to obtain blood samples, which brings inconvenience and discomfort to diabetic patients. Sweat glucose content has been proven to be related to blood glucose concentration, so sweat glucose detection shows significant advantages in non-invasive blood glucose monitoring. The performance of enzyme-based glucose sensors is easily affected by external environments such as temperature and pH. Therefore, there is an urgent need to develop non-enzymatic glucose sensors with good stability, low cost and high performance for sweat glucose detection.
Results:
In this study, a non-enzymatic glucose sensor had been successfully fabricated by integrating copper oxide (CuO) nanoparticles with tungsten disulfide (WS2) nanosheets obtained via a liquid-phase exfoliation method. The sensor was characterized using scanning electron microscopy (SEM) and electrochemical measurements, showing that WS2/CuO nanocomposites formed a honeycombed-like surface improving the catalytic ability of glucose. The fabricated non-enzymatic amperometric glucose sensor is capable of detecting a wide range of target concentrations, from 0.1 mM to 20 mM. The experiment found that the sensor response signal has a linear relationship with the target concentration in the range of 0.1 mM-2 mM and 2 mM-20 mM, respectively, and the corresponding detection sensitivities are 311.23 μA mM-1 cm-2 and 238.3 μA mM-1 cm-2 under optimized experimental conditions. The sensor has a detection limit as low as 0.65 μM. The sensor also has good selectivity, repeatability, reproducibility and stability. Finally, the fabricated sensor was applied to the detection of glucose in artificial sweat and human sweat.
Significance:
The experimental results compared with the spectrophotometric method illustrate the excellent accuracy of the non-enzymatic glucose sensor proposed in this study in detecting real sweat sample analysis, highlighting its great potential for future practical applications.
More Related Videos
08:19Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Related Concept Videos
Amperometry: Overview
Accessory Structures of the Skin: Sweat Glands
Sweat glands are classified as merocrine glands; that is, the secretions are excreted by exocytosis through a duct without affecting the cells of the gland. There...