The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Left Atrial Appendage Occlusion Using a Novel System: Preclinical Evaluation and First-in-Human Experience
Felix Mahfoud1, Christian Ukena2, Philip D Adamson3
1Department of Cardiology, University Heart Center, University Hospital Basel, Basel, Switzerland; Cardiovascular Research Institute Basel, University Heart Center, University Hospital Basel, Basel, Switzerland; Department of Biomedical Engineering, University Basel, Basel, Switzerland.
Background:
Atrial fibrillation is a major stroke risk, with >90% of thrombi originating from the left atrial appendage (LAA). Oral anticoagulation is effective, but many patients cannot use it long term. Current percutaneous LAA occlusion devices reduce stroke risk but are limited by device-related thrombus (DRT) and peridevice leak.
Objectives:
The aim of this study was to assess the safety and performance of the novel Zenith LAA occlusion system in both preclinical canine models and a prospective first-in-human (FIH) clinical study.
Methods:
A multistage evaluation included 3 preclinical trials in 17 canines (≤90-day follow-up) and a single-arm, prospective FIH study in 10 nonvalvular atrial fibrillation patients. This novel LAA device consists of a conformable Nitinol scaffold covered with a polyethylene terephthalate membrane, a self-closing central conduit to minimize exposed metal, and an independently deployed dynamic anchoring system. Primary imaging follow-up was 45-day transesophageal echocardiography.
Results:
In a preclinical canine model (n = 17), explant analysis at 45 days demonstrated complete endothelialization with no evidence of DRT or significant inflammation. In the FIH study, 9 of 10 patients (90%) underwent successful implantation. The a priori primary safety endpoint of acute procedural success was achieved in 9 of 9 patients (100%), and the a priori primary performance endpoint of mechanical device closure at 45 days was also achieved in 9 of 9 patients (100%). At 45-day transesophageal echocardiography, all implants were stable, with complete protocol-defined LAA closure and no DRT. No device- or procedure-related serious adverse events occurred.
Conclusions:
This novel LAA occlusion system demonstrated favorable early safety and performance in both preclinical and FIH evaluation. The novel design features enabled effective LAA sealing and absence of DRT, supporting further investigation in larger trials.
More Related Videos
Related Concept Videos
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Preclinical Development: Overview
First Pass Effect
Depending on the route of administration, drugs can be metabolized in the liver, intestine, lungs, and vasculature. Orally administered drugs are first absorbed through the...
Assessment of the Gastrointestinal System I: Subjective Data
Health History
The initial step in assessing the GI system is obtaining a comprehensive health history. This includes inquiring about the patient's history or presence of problems...
Bioavailability Study Design: Healthy Subjects Versus Patients