Related Experiment Video
Updated: Jan 14, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Long-Term Implantable Cardioverter Defibrillator Lead Dysfunction After Left Ventricular Assist Device Implantation
Tsukasa Oshima1, Kohei Ishibashi1, Kenichiro Yamagata2
1Department of Cardiovascular Medicine, National Cerebral and Cardiovascular Center, Suita, Japan.
Background:
Implantable cardioverter defibrillator (ICD) lead dysfunction after left ventricular assist device (LVAD) implantation can occur, but only short-term outcomes have been reported.
Objectives:
We aimed to evaluate the long-term incidence, characteristics, and predictors of persistent ICD lead dysfunction after LVAD implantation.
Methods:
This was a retrospective multicenter study. All patients with a transvenous ICD lead at the time of LVAD implantation between January 1, 2011, and December 31, 2023, were enrolled. The primary endpoint was lead dysfunction. Risk factors for persistent compared to temporary lead dysfunction using a logistic analysis were determined.
Results:
A total of 170 patients (mean age: 48.0 ± 12.7 years) were analyzed. The median follow-up period was 46.2 (Q1-Q3: 32.3-61.4) months. Lead dysfunction was observed in 124 leads (72.9%), of which 60.4% (N = 75) occurred within a year after LVAD implantation. Of the 124 lead dysfunctions, 84 (67.7%) showed persistent dysfunction and 40 (32.3%) were temporary. Lead dysfunction occurring later than 2.1 months after LVAD implantation was a risk factor for persistent lead dysfunction (area under the curve: 0.673). Multivariable analysis identified lead dysfunction occurring later than 2.1 months after LVAD implantation as an independent risk factor for persistent lead dysfunction (OR: 4.67; 95% CI: 2.05-10.94; P < 0.001).
Conclusions:
In patients undergoing LVAD implantation, ICD lead dysfunction was observed in 72.9% after a mean follow-up of 46.2 months, of which two-thirds had persistence lead dysfunction. Later occurrence of lead dysfunction (>2 months after implant) may be a risk factor for persistent lead dysfunction.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management