The Global Outbreak of M. chimaera Infection Following Cardiac Surgery: Another Piece of the Puzzle

Savina Ditommaso1, Gabriele Memoli1, Francesca Anselmi2

  • 1Department of Public Health and Paediatrics, University of Turin, 10124 Torino, TO, Italy.

PubMed

Insights

Mycobacterium chimaera infections linked to heater-cooler units (HCUs) were investigated. Whole-genome sequencing confirmed strains from Italian HCUs match those in global outbreaks, suggesting production site contamination.

Area of Science:

  • Medical Microbiology
  • Infectious Diseases
  • Genomics

Background:

  • Invasive Mycobacterium chimaera infections have been reported globally, primarily associated with heater-cooler units (HCUs) used in cardiac surgery.
  • These infections are often linked to aerosolized water from HCU tanks, posing a risk to patients undergoing extracorporeal circulation.
  • An outbreak of five M. chimaera cases linked to Stockert 3T (LivaNova) HCUs occurred in Italy's Piedmont region.

Purpose of the Study:

  • To investigate the genetic relatedness of M. chimaera strains isolated from HCUs in the Piedmont region to those implicated in global outbreaks.
  • To determine the prevalence of M. chimaera contamination in Stockert 3T HCUs within regional cardiac surgery operating rooms.

Main Methods:

  • Whole-genome sequencing (WGS) was performed on M. chimaera strains isolated from 10 Stockert 3T HCUs.
  • M. chimaera isolates were analyzed according to van Ingen's criteria for classification.
  • Prevalence of M. chimaera in monitored HCUs was assessed.

Main Results:

  • Mycobacterium chimaera was detected in 59% of the monitored Stockert 3T HCUs.
  • WGS analysis revealed that four of the ten sequenced M. chimaera strains belonged to subgroups 1.1 or 1.8.
  • These subgroups are known to be associated with the worldwide HCU-related M. chimaera outbreaks.

Conclusions:

  • The findings provide further evidence supporting the hypothesis that M. chimaera contamination originates at the LivaNova production site.
  • Exposure to contaminated factory plumbing systems is implicated as the source of M. chimaera in the HCUs.
  • This highlights the need for stringent quality control measures during HCU manufacturing to prevent patient infections.