Related Experiment Video
Updated: Jan 12, 2026

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Cylindrical Lead Spinal Cord Stimulators: Common Device-Related Issues
Jamal Hasoon1, Omar Viswanath2,3, Thomas T Simopoulos4
11. Department of Anesthesiology, Critical Care, and Pain Medicine University of Texas Health Science Center at Houston.
Abstract:
A comprehensive review of device-related complications from cylindrical lead spinal cord stimulator (SCS) implants was previously published. [1] Although SCS remains a minimally invasive treatment for neuropathic pain, hardware-related problems such as lead migration, lead fracture, lead disconnection, generator malfunction, loss of charge, generator flipping, pocket pain, and paresthesia intolerance are reported more often than biologic complications. [1,2] Understanding these issues, their clinical presentations, diagnostic evaluation, and management strategies is critical for long-term outcomes. This infographic was designed to present the key findings of the review in a clear and accessible visual format.