Feasibility study of a multi-lesion cyberknife radiotherapy plan verification method using a 2D array with pre-set
Tengxiang Li1,2, Jinhu Chen1, Ruimin He2
1Department of Radiation Physics Shandong Cancer Hospital and Institute Shandong First Medical University and Shandong Academy of Medical Sciences Jinan Shandong Province P. R. China.
A new method using SRSmapcheck with pre-set roll angles validates intracranial multi-lesion CyberKnife plans, measuring more targets per plan. This approach enhances dose distribution accuracy for complex treatments.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiosurgery
Background:
- Validating intracranial multi-lesion CyberKnife plans with SRSmapcheck is challenging due to limitations in measuring multiple targets and lack of roll verification.
- Existing methods using a fixed 0° array setting only measure one target, leaving others unverified.
- The CyberKnife treatment planning system lacks support for roll verification tools, and X-sight fiducial marker guidance is incompatible with free array roll.
Purpose of the Study:
- To develop and validate a novel method for verifying intracranial multi-lesion CyberKnife plans using SRSmapcheck with pre-set roll angles.
- To optimize dose distribution verification in high-dose and rapidly varying areas for multi-lesion radiosurgery.
- To compare the efficacy of the pre-set roll angle method against the traditional fixed 0° verification method.
Main Methods:
- A geometric model was created to determine the relationship between SRSmapcheck array roll angle and tumor spacing.
- SRSmapcheck and StereoPHAN computed tomography (CT) images were acquired at 15° intervals for phantom plans.
- 101 intracranial multi-lesion plans were verified using both fixed 0° and pre-set roll angle methods, with subsequent comparative analysis.
Main Results:
- The pre-set roll angle method assessed an average of 2.34 targets per plan, compared to 1.32 targets with the fixed 0° method.
- Plan pass rates remained stable between methods for gamma analysis criteria of 2 mm/1% and 2 mm/2%.
- Recommended gamma analysis criteria for pre-set roll angle verification are 1 mm/1% and 1 mm/2% for optimal results.
Conclusions:
- SRSmapcheck with the pre-set roll angle method enables comprehensive validation of intracranial multi-lesion CyberKnife plans.
- This novel approach allows for the measurement of multiple targets within a single validation session.
- The study recommends specific gamma analysis criteria (1 mm/1% and 1 mm/2%) for effective pre-set roll angle verification.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias