Related Experiment Video
Updated: Jan 12, 2026

A Quantitative Glycomics and Proteomics Combined Purification Strategy
Published on: March 8, 2016
Development of a fit-for-purpose semi-homogeneous competitive ELISA for total PEG quantitation in a PEGylated
Nancy Yu1, Vincent Chang1,2, Ihsan Nijem1
1BioAnalytical Sciences, Genentech, South San Francisco, CA, USA.
Aim:
The therapeutic effectiveness of a drug could be significantly enhanced by prolonging its retention time. Polyethylene glycosylation (PEGylation) technology is a method to increase the half-life of protein therapeutics. PEGylated anti-Factor D Fab (PEG-aFD) was developed as a potential therapeutic, targeting factor D to reduce the frequency of intravitreal injections in patients with geographic atrophy. To support a GLP Toxicology study, there was a need to develop a PEG assay to measure total PEG concentration.
Result:
PEG-aFD consists of Fabs conjugated to a multi-arm PEG molecule, which posed challenges for PEG assay development. A sensitive fit-for-purpose, semi-homogeneous competitive ELISA (Enzyme-Linked Immunosorbent Assay) using a biotinylated version of the therapeutic as the competing molecule was successfully developed and validated with LLOQ (lower limit of quantification) at 200 ng/mL. The assay was used to measure total PEG concentrations in cynomolgus monkey serum samples from the study. The results revealed measurable PEG levels (ranging from 0.468 μg/mL to 188 µg/mL) in all samples across all dosage levels, but no PEG-related toxicity was observed in all the animals.
Conclusion:
A competitive ELISA was successfully developed and validated to measure total PEG concentrations in cynomolgus monkey serum samples to support a GLP Toxicology study.
More Related Videos
08:28Validated Immunochemical Assay for Comprehensive Determination of the Human Epidermal Growth Factor Receptor 2 Released from and Bound to Cells
Published on: May 9, 2025
09:32Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017