Impact of Pre-Actuation Shaking on the Aerosolization Performance of Easyhaler® Salbutamol Dry Powder Inhaler
Ludovica Esposito1,2, Patrick He2, Chun Yuen Jerry Wong2,3
1Department of Biomolecular Sciences, University of Urbino Carlo Bo, Urbino, PU, Italy.
Abstract:
PURPOSE OR OBJECTIVE: Each Easyhaler® Salbutamol 100 μg dry powder inhaler (DPI) uniquely requires pre-actuation shaking, unlike most commercial DPIs. This study aimed to investigate the aerodynamic and functional implications of pre-actuation shaking to determine its effect on drug delivery performance.
Methods:
Device actuation was performed following 0 to 5 pre-actuation shakes. Performance was assessed using a multi-technique approach: cascade impaction, dose uniformity testing (DUSA), laser diffraction (SprayTec), optical microscopy, and scanning electron microscopy (SEM). Key parameters measured included emitted dose (ED), delivered dose (DD), fine particle dose (FPD), fine particle fraction (FPF), mass median aerodynamic diameter (MMAD), and throat deposition.
Results:
Delivered dose remained relatively constant regardless of shaking. However, FPD and FPF significantly improved with increased shaking, particularly at three to five shakes. Cascade impaction demonstrated reduced throat deposition and greater deposition in stages 3-4 (< 5 µm respirable fraction) under these conditions. An inverse correlation between throat deposition and FPF was identified. SEM and microscopy confirmed consistent particle morphology and blend uniformity, while laser diffraction showed a predominance of larger carrier particles under 0-shake conditions.
Conclusions:
Pre-actuation shaking substantially influences the aerosolization performance of Easyhaler® Salbutamol. At least three vertical shakes are required to achieve optimal fine particle delivery and reduce throat deposition. These findings highlight the importance of proper patient instruction on device handling. Further studies should assess whether similar effects occur with other Easyhaler® formulations.
Related Concept Videos
Inhaled Medications
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Antiasthma Drugs: β2-Adrenoceptor Agonists
One class of bronchodilators includes β2-adrenoceptor agonists. These agents target the β2-adrenoceptors located on bronchial smooth muscle cells. By stimulating these receptors, β2-agonists induce...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Adrenergic Agonists: Mixed-Action Agents
Ephedrine and pseudoephedrine lack a catecholamine group, making them less susceptible to degradation by metabolic enzymes. They have increased oral bioavailability and lipophilicity, resulting in a longer duration of action. Their response is reduced by...
Upper Respiratory Drugs: Antitussives, Expectorants, and Mucolytics
Antitussives include codeine, dextromethorphan (Robitussin), and benzonatate (Tessalon). Codeine and dextromethorphan exert their effects centrally by suppressing the cough reflex center in the medulla. Benzonatate operates peripherally within the respiratory tract by...