Device-based day-to-day and observer variability to quantify dilation capacity in the retinal microcirculation
Lukas Streese1,2, Christoph Hauser1, Denis Infanger1
1Department of Sport, Exercise and Health, Medical Faculty, University of Basel, Basel, Switzerland.
Frontiers in Physiology
|October 31, 2025
Summary
Dynamic retinal analysis (DVA) shows moderate day-to-day variability but excellent intra- and interobserver reliability. Adjusting for baseline diameter may improve DVA
Area of Science:
- Ophthalmology and visual sciences
- Medical technology and devices
- Cardiovascular research
Background:
- Dynamic retinal analysis (DVA) is a validated method for quantifying microvascular endothelial function.
- Assessing the reliability of DVA is crucial for its clinical implementation.
- Understanding variability is key to standardizing DVA protocols.
Purpose of the Study:
- To evaluate day-to-day, intra- and interobserver variability of DVA.
- To compare DVA measurements between two device generations (DVA 2.0 and DVA 3.0).
- To identify factors influencing DVA reliability.
Main Methods:
- DVA measurements were taken on two separate days using DVA 2.0 and DVA 3.0 devices.
- Intraobserver variability was assessed by one reader analyzing 20 arteriolar (aFID) and venular flicker-light induced dilation (vFID) signals twice.
- Interobserver variability was assessed by a second reader independently analyzing 20 aFID and vFID signals. Interclass correlation coefficient (ICC) was used.
Main Results:
- Moderate to good day-to-day variability was observed for both aFID and vFID across both DVA versions.
- Excellent intraobserver (aFID/vFID: 0.999) and interobserver (aFID: 0.997, vFID: 0.998) reliability was demonstrated.
- Moderate interdevice variability was found between DVA 2.0 and DVA 3.0 (aFID: 0.76, vFID: 0.87). ICC improved after correcting for baseline diameter.
Conclusions:
- DVA exhibits reliable intra- and interobserver performance.
- Day-to-day and interdevice variability exist, suggesting a need for standardization.
- Accounting for arteriolar baseline diameter variations can enhance DVA reliability in research and clinical settings.