Comparison of precise imaging and iterative reconstruction techniques at low doses using the dose right index in
Imnejongla Chang1, Abhimanyu Pradhan1, Sharath S1
1Department of Medical Imaging Technology, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, India.
Background:
Artificial intelligence (AI)-driven image reconstruction has shown potential for enhancing image quality in CT angiography while reducing radiation exposure. However, reproducibility and methodological transparency are essential for real-world adoption.
Purpose:
To compare image quality and radiation dose of cerebral CT angiography reconstructed with conventional iterative reconstruction (IR) and AI-based Precise Imaging (PI) at 100 kVp using a reduced Dose Right Index (DRI-15).
Methods:
In this prospective single-center study, 68 patients underwent 100 kVp cerebral CT angiography with a 50% reduced Dose Right Index (DRI-15 vs. DRI-36). Identical raw datasets were reconstructed using IR (iDose4) and PI. Quantitative analysis included attenuation, image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR), while two blinded radiologists independently performed qualitative assessment using a 5-point Likert scale. Radiation dose metrics (CTDIvol, DLP, effective dose) were recorded. Standardized acquisition protocols, fixed contrast injection, and consistent ROI placement ensured reproducibility.
Results:
PI reconstructions yielded significantly lower noise and higher SNR and CNR than IR (all p < 0.001; r = 0.61-0.71), with no difference in vascular attenuation (p = 0.118). Qualitative scores were superior for PI across all criteria (p < 0.01), while inter-rater agreement was substantial for both methods, slightly higher with IR. The median effective dose was 0.785 mSv (IQR 0.71-0.90).
Conclusion:
AI-based PI reconstruction enhances image quality without increasing radiation dose or altering vascular attenuation, consistently outperforming IR and supporting its use in low-dose cerebral CT angiography, especially in dose-sensitive patients.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies VII: Vascular Imaging