Related Experiment Video
Updated: Jul 26, 2026

Training Persons with Spinal Cord Injury to Ambulate Using a Powered Exoskeleton
Published on: June 16, 2016
Influence of gait-synchronized functional electrical stimulation during exoskeleton-assisted ambulation on
Robert Voicu1, Daniela B Kuchen2, Claudio Perret3
1Neuro-Musculoskeletal Functioning and Mobility Group, Swiss Paraplegic Research, Nottwil, Switzerland; Department of Health Sciences and Technology, ETH Zurich, Zurich, Switzerland.
Objective:
To evaluate the impact of gait-synchronized functional electrical stimulation during exoskeleton-assisted ambulation on cardiorespiratory demand in individuals with incomplete spinal cord injury.
Design:
Cross-sectional study employing a randomized crossover arrangement of measurements.
Participants:
Convenience sample of 11 individuals with chronic incomplete spinal cord injury and partial walking ability.
Methods:
Participants completed 2 x 6-min walking tests (6MWTs) with the EksoNR, 1 with and 1 with-out gait-synchronized functional electrical stimulation targeting gait-related muscles in a randomized order. Cardiorespiratory and metabolic parameters were measured breath-by-breath via ergospirometry. The primary outcome was oxygen consumption (V̇O2/kg). Secondary outcomes included further cardiovascular and metabolic parameters. Data from the final 2 min of each 6MWT were analysed using linear mixed-effect models.
Results:
V̇O2/kg increased by 6% with functional electrical stimulation compared with exoskeleton-assisted ambulation alone (15.07 ± 4.11 vs 14.21 ± 3.61 mL·min-¹·kg-¹, p = 0.02). Heart rate, ventilation, and energy expenditure were also elevated (p < 0.05), while breathing frequency, respiratory exchange ratio, distance, and perceived exertion remained unchanged. Inter-individual differences in V̇O2/kg were not explained by stimulation amplitude (r = 0.36, p = 0.27).
Conclusion:
Adding functional electrical stimulation to exoskeleton-assisted gait therapy consistently increased cardiorespiratory demand, potentially enhancing training intensity. Further research assessing long-term clinical impact is required.