Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Does the ratio of stem-to-femur length matter? A validated finite element analysis
Georgios Banis1, Charoula Kousiatza1, Maria Sarkiri1
1Akmelogi SMPC Athens Greece.
Purpose:
Total hip arthroplasty is among the most successful orthopedic procedures; however, approximately 10%-20% of patients require revision surgery within 15-20 years. Potential failure mechanisms at the bone-implant interface include stress shielding and excessive micromotion. This experimental work aims at examining the optimal ratio of femur length and stem length with respect to primary stability and stress shielding effect, employing a validated finite element analysis.
Methods:
One hundred human femora were reconstructed from computed tomography (CT) scans, and for each femur, 13 personalized cementless stems were designed with varying lengths (70-130 mm, in 5 mm increments), resulting in 1300 finite element models. Each model was analyzed under standardized joint and muscle loading conditions. The model was validated experimentally after cadaveric implantation with digital image correlation (DIC) under static loading. Primary outcomes included micromotion at the bone-implant interface and proximal stress shielding. Optimal stem length was determined as the ratio of stem-to-femur length that minimized the combined normalized scores of micromotion and stress shielding. The effect of stem-to-femur length ratio was analysed with linear regression.
Results:
The analysis revealed a correlation between femoral length and stem length. Longer stems provided better initial stability and reduced micromotion. On the other hand, shorter stems demonstrated less proximal stress shielding and a better stress distribution resembling that of the intact femur. The optimal stem length was found to be 18.5% ± 2.7% (mean ± SD) of the femur length in terms of the mean of calculated micromotions and stress shielding, indicating that personalized stem length selection can enhance the bone-implant osseointegration.
Conclusion:
The selection of a stem length that corresponds proportionally to femoral length can improve primary stability and stress shielding upon implantation, highlighting the importance of enhancing initial fixation in order to extend the lifespan of hip replacements.
Level Of Evidence:
N/A.
Related Concept Videos
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...
Euler's Formula to Columns with Other End Conditions
Indeterminate Structure
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Design of Columns under a Centric Load
Euler's formula is applicable under the assumption that the column is a perfect, straight, homogenous prism, and it is operating...
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to its distance...