Open-Source Simulator of Imaging Near Metal at Arbitrary Magnetic Field Strengths
Kübra Keskin1, Ana R Sanson1, Brian A Hargreaves2,3,4
1Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
Purpose:
The increasing prevalence of orthopedic metallic implants necessitates optimization of MRI methods to monitor surrounding tissues and identify complications. The emergence of contemporary mid- and low-field systems as promising platforms for imaging around metal requires development of new protocols optimized for these field strengths. Motivated by these, we develop a simulation framework that predicts MRI performance near metallic implants, considering B0 field strength and multi-spectral imaging (MSI) parameters.
Methods:
A simulation framework for imaging near metal incorporating 3D implant and 3D human anatomical models, and imaging and sequence parameters is developed. MSI acquisitions are simulated based on susceptibility-induced field shifts. Validation is performed using phantom and in vivo experiments at 0.55 and 3 T, comparing simulated and experimental images for artifact shape and SNR. Performance is demonstrated by evaluating impacts of field strength, spectral bins, RF and readout bandwidths, and implant materials.
Results:
Simulations accurately reproduced experimental imaging performance, with extent and patterns of metal artifacts closely matching experimental observations at both 0.55 and 3 T, while correctly predicting increased SNR and artifact size with increasing field strength. The framework demonstrated the impact of field strength, spectral bins, RF and readout bandwidths, and implant material on metal artifacts and image quality.
Conclusion:
This open-source simulation framework effectively predicts MRI performance near metallic implants. This can facilitate protocol optimization and exploration of imaging parameters without extensive/costly in vivo experiments. This tool can enable researchers to predict imaging performance for various implants across different field strengths, potentially accelerating development of optimized protocols for clinical use.
Related Concept Videos
Magnetic Field Of A Current Loop
Magnetic Resonance Imaging
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...
Magnetic Field Due To A Thin Straight Wire
Magnetic Field due to Moving Charges
Consider a point charge moving with a constant velocity. Like the electric field, the magnetic field at any point is directly proportional to the magnitude of the charge and inversely proportional to the square of the distance between the source point and the field point. However, unlike the electric field, the magnetic field is always perpendicular to the plane containing the line...
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...