Related Experiment Video
Updated: Jun 24, 2026

Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
A hybrid IGRT workflow using SGRT and CBCT for prostate SBRT: Feasibility, efficiency, and safety
Juergen Meyer1, Angelia Tran1, Ting Martin Ma1
1Department of Radiation Oncology, Fred Hutchinson Cancer Center, University of Washington, Seattle, Washington, USA.
Background And Purpose:
Safe delivery of prostate stereotactic body radiotherapy (SBRT) relies on precise target localization. Without access to real-time intrafraction motion management, careful optimization of IGRT protocols is necessary to safeguard treatment accuracy and patient outcomes.
Methods:
An IGRT workflow is proposed that incorporates surface-monitoring (SGRT) to complement cone-beam CT (CBCT) imaging. The study evaluates 23 consecutive SBRT prostate patients who were treated on a prospective registry study. Each patient received pre- and mid-treatment and a subset received post-treatment CBCTs. The frequency and magnitude of SGRT triggered beam interruptions as well as treatment times were recorded.
Results:
The median number of CBCTs acquired per fraction was four and the median treatment time was 23 min (IQR 19-27). SGRT detected intra-fraction surface-based motion beyond a combined 4 mm vector isocenter tolerance in 62% of all fractions treated, with a maximum motion of 15 mm. On average < 2 beam interruptions were triggered by SGRT per treatment fraction. There was no statistically significant correlation between overall treatment time and SGRT-triggered beam interruptions (r = 0.048, p = 0.645). There was a weak but statistically relevant correlation of overall treatment time with the maximum detected motion (r = 0.23, p = 0.026). SGRT detected five fractions where the patients had persistently moved outside the SGRT tolerance, and for three of these (60%), a CBCT verified that the target was out of tolerance.
Conclusion:
SGRT is a valuable tool that complements CBCT-based IGRT. An SGRT motion vector tolerance of 4 mm provides a pragmatic compromise between detecting patient motion and treatment efficiency. Overall, persistent patient motion during treatment was infrequent in this cohort, however, SGRT was able to detect several cases where the internal target was outside of the tolerance highlighting that patient monitoring with SGRT can contribute to improved quality and safety for prostate SBRT.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography