Related Experiment Video
Updated: Jan 12, 2026

Author Spotlight: Expanding Interventional Pulmonology Research with Robotic-Assisted Bronchoscopy
Published on: July 19, 2024
First European Results of Shape-Sensing Robotic-Assisted Bronchoscopy
Judith Maria Brock1,2, A Susanne Dittrich3,4, Konstantina Kontogianni3,4
1Department of Pneumology and Critical Care Medicine, Thoraxklinik, University of Heidelberg, Heidelberg, Germany, judith.brock@med.uni-heidelberg.de.
Introduction:
Shape-sensing robotic-assisted bronchoscopy (ssRAB) was shown to be superior in diagnosing peripheral pulmonary nodules (PPNs) compared to conventional bronchoscopy. Although RAB is established in the USA since 2019, the ION™ Endoluminal System was not CE marked in Europe until 2023.
Methods:
The first prospective European single-center, premarket study, conducted between 2022 and 2023, assessed the diagnosis of PPNs with ssRAB, using the ION™ Endoluminal System. Patients with suspected lung cancer or metastasis, PPNs of 1-3 cm, and ≥3 airway generations out were included in this study and followed up for up to 6 months. The primary outcome was the rate of tool-in-lesion (TIL), confirmed by mobile cone-beam computed tomography or with malignant index biopsy finding. Secondary outcomes included procedural characteristics, diagnostic yield, and adverse events.
Results:
A total of 43 patients with a mean nodule size of 19.2 × 16.9 × 15.8 mm, a mean distance of 17.4 ± 15.5 mm to the pleura, and mean 6.6th generation of airway were analyzed. TIL was achieved in 90.7% of all cases. Strict diagnostic yield was 67.4% and sensitivity for malignancy was 78.6%. No pneumothorax and no adverse events were reported outside of 3 cases of Nashville ≤2 bleeding. Predictors for successful diagnosis were the inner or middle third location (OR: 4.19, p = 0.039), CT bronchus sign (OR: 4, p = 0.044), and distance from pleural wall (OR: 1.05, p = 0.048). The lower lobe location (OR: 0.06, p < 0.001) was associated with nondiagnostic cases.
Conclusion:
The first European cases show ssRAB is a safe procedure with promising results for enabling diagnosis of PPNs.
Related Concept Videos
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques