Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Validation of heart rate measured by the consumer-level wearable device Fitbit: a cross-sectional observational study
Daisuke Higuchi1, Yuko Takahashi1, Shigeya Tanaka1
1Department of Physical Therapy, Faculty of Health Care, Takasaki University of Health and Welfare: 27 Nakaorui-machi, Takasaki-shi, Gunma 370-0033, Japan.
Abstract:
[Purpose] The present study aimed to validate the heart rate of the consumer-level wearable device Fitbit Inspire 3. [Participants and Methods] Participants were 22 healthy university students (20.9 ± 0.6 years, 11 males and 11 females). They performed 5-minute resting lying, sitting, walking (3.2 km/h, 4.2 km/h, 6.0 km/h and 3.2 km/h with 3 kg backpack) and running (8.4 km/h running) on a treadmill, and stair climbing, with Fitbit- and electrocardiogram-based heart rate (Fitbit-HR and ECG-HR). The relationship between Fitbit-HR and ECG-HR was evaluated using a generalized linear mixed model. Additionally, Brandt-Altman analysis and absolute percentage error were used to examine the measurement error. [Results] Fitbit-HR was strongly associated with ECG-HR and oxygen uptake. In the Brandt-Altman analysis with Fitbit-HR and reference ECG-HR, most plots were within the limit of agreement. Fitbit-HR showed significant fixed errors, which were, on average, three beats smaller than those of ECG-HR. No significant proportional errors are observed. The mean absolute percentage error for most of the tasks was less than 10%. [Conclusion] The Fitbit Inspire 3 showed a strong correlation and clinically acceptable agreement with ECG-HR in healthy young adults, despite a small, consistent underestimation. Its validity was supported by its association with oxygen uptake.
Related Concept Videos
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring pulse
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: