Differences in Impulse Oscillometry Parameters Measured Using Different Devices
Chalerm Liwsrisakun1, Athavudh Deesomchok1, Chaicharn Pothirat1
1Profs. Liwsrisakun, Deesomchok, Pothirat, Ms. Duangjit, and Dr. Chaiwong are affiliated with Division of Pulmonary, Critical Care, and Allergy, Department of Internal Medicine, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand.
Impulse oscillometry (IOS) measurements of reactance, resonant frequency (Fres), and area under the reactance curve (AX) differ significantly between MasterScreen-IOS and MostGraph-02 devices. These differences mean IOS results are not interchangeable for tracking lung disease progression.
Area of Science:
- Pulmonary Function Testing
- Respiratory Physiology
- Medical Device Comparison
Background:
- Limited data exist on absolute and percent predicted impulse oscillometry (IOS) values across different devices.
- Understanding device-specific IOS performance is crucial for accurate interpretation in chronic respiratory diseases.
Purpose of the Study:
- To compare impulse oscillometry (IOS) values and percent predicted values between MasterScreen-IOS and MostGraph-02 devices.
- To assess the agreement and correlation of IOS measurements obtained from two different devices in patients with COPD and asthma.
Main Methods:
- Retrospective analysis of IOS data from 54 subjects (asthma and COPD) using both MasterScreen-IOS and MostGraph-02.
- Paired sample t-tests and Pearson's correlation were used for comparing IOS values.
- Bland-Altman analysis determined the limits of agreement between the two devices.
Main Results:
- Significant differences were observed in absolute and percent predicted reactance values, including resonant frequency (Fres) and area under the reactance curve (AX), between the two devices.
- MasterScreen-IOS yielded higher reactance values compared to MostGraph-02.
- Discrepancies were noted in percent predicted values for resistance heterogeneity (R5-R20), Fres, and AX.
Conclusions:
- Significant differences in IOS reactance values exist between MasterScreen-IOS and MostGraph-02 devices.
- Device-specific reference equations likely contribute to variations in percent predicted values.
- IOS measurements from these devices should be interpreted cautiously and are not interchangeable for longitudinal assessments.
Related Concept Videos
Special considerations while measuring pulse
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Voltammetry: Factors Affecting Measurements
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...