Related Experiment Video
Updated: Jan 11, 2026

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohort
Peggy Jacon1, Mouna Benkilani2, Arnaud Bisson3
1Department of Cardiology, University Hospital of Grenoble Alpes, 38043 Grenoble, France.
Aims:
Leadless pacemakers (LLPs) offer a valuable option for selected patients requiring ventricular pacing. However, data on revision procedures-defined as any intervention on an implanted LLP to correct dysfunction or upgrade the system, with or without removal-remain scarce. This study aimed to evaluate the real-world incidence, characteristics, and outcomes of LLP revisions, with particular emphasis on the feasibility and safety of removal.
Methods And Results:
This multicentre retrospective study involved 46 French centres performing LLP implantations between 2015 and 2023. Among 8994 Micra LLPs implanted, 100 revision procedures (1.1%) were performed in 100 patients, including 73 Micra VR and 27 Micra AV devices, at a median follow-up of 229 (68.5-629.8) days. Seventy revisions were managed without device removal, while 30 involved removal. Revision indications included device upgrade (55%), high pacing threshold (33%), battery depletion (3%), ventricular arrhythmias (3%), and miscellaneous causes (6%). No significant differences were observed in pacing indications, implantation characteristics, or initial device indication between the two groups. Revisions occurred significantly earlier in the removal group than in the group without removal [median 81.5 (8-211.5) days, vs. 334 (130.25-882.5) days, P < 0.001], with 80% occurring within 1 year. Device removal was successful in 97% of cases (one failure at 157 days of follow-up), with no major complications reported.
Conclusion:
LLP revisions are infrequent and predominantly managed without device removal. However, given the high procedural success rate and the low complication rate, device removal appears to be underutilized and should be considered more frequently on a case-by-case basis.
More Related Videos
Related Concept Videos
Mitral Regurgitation III: Medical Management
Rheumatic Heart Disease II: Clinical Manifestations and Diagnostic Studies
Mitral Stenosis III: Medical Management
Mitral Regurgitation I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Rheumatic Heart Disease III: Medical Management