Related Experiment Video
Updated: Jan 11, 2026

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Analytical Performance of Adams A1c HA-8180V: Model for Quality Targets Versus Biological Variation and Sigma Metrics
Background:
HbA1c is a biomarker for the diagnosis of type 2 diabetes (T2D) and prediabetes, in addition to its role in monitoring glycemic control. These indications require mastery of analytical performance. This study evaluated the analytical performance of the ADAMS A1C HA-8180V analyzer used in the biochemistry laboratory of Mohammed V Military Hospital in Rabat (Morocco).
Methods:
This is a retrospective study conducted between April 2023 and April 2024. The results of daily internal quality control (IQC) and monthly external quality assessment (EQA) were used to calculate imprecision and accuracy. Three analytical-mathematical models based on the total acceptable error (TAE) were used: biological variation (BV), Sigma metrics (SM), and Model for Quality Targets (MQT) recommended by the IFCC (International Federation of Clinical Chemistry and Laboratory Medicine).
Results:
The mean CVs of the IQC were 0.82% and 0.92% for levels 1 and 2, respectively, significantly lower than the 3.4% CV recommended by the state of the art. The mean relative bias of the EQA was 1.47%. According to the BV model, the "desirable" performance level predominated. According to the SM model, the majority of results exceeded 5 sigma. Applying the MQT model, all results exceeded the 2 sigma recommended by the IFCC.
Conclusions:
The HbA1c results reported by the laboratory met the IFCC criteria for routine laboratories provid-ing T2D diagnosis and monitoring of diabetic patients.
More Related Videos
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Statistical Analysis: Overview
One of the most commonly used statistical quantifiers is the mean, which is the ratio between the sum of the numerical values of all results and the...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Interpreting X̄ Charts
An x̄ chart plots the values of individual measurements over time against control limits calculated from historical data. The central line...
Data Validation
Key parameters for method validation include:
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...